OPERATING INSTRUCTIONS
SENSO2® OXYGEN PURITY INDICATOR- IF YOUR UNIT FEATURES THE O2 SENSOR FEATURE
This feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate.
INITIAL STARTUP OF THE CONCENTRATOR
NOTE: Concentrator may be used during the initial start
When the unit is initially turned on, the GREEN power LED (I/O) will illuminate. Within 5 to 30 minutes after initial startup, a GREEN light will illuminate in the Oxygen Purity section of the Control Panel (FIGURE
7)if concentration is within specification. Any other SensO2® indicator lights illuminating will signal a condition that requires a call to the supplier.
EXPLANATION OF OXYGEN PURITY INDICATOR LIGHTS (FIGURE 7):
GREEN LIGHT ( O2 ) - Normal Operation
YELLOW LIGHT ( ) - Call Supplier IMMEDIATELY. You may continue to use the concentrator unless instructed otherwise by your supplier. Be certain that BACKUP OXYGEN is nearby.
RED LIGHT ( )- Total unit shutdown. Switch IMMEDIATELY to BACKUP OXYGEN supply. Call Supplier IMMEDIATELY.
GREEN LIGHT - with YELLOW LIGHT FLASHING - Call Supplier IMMEDIATELY. Oxygen sensor malfunctioning; you may continue to use the concentrator.
Oxygen Purity
Auto Shut - Down | RED |
Use Backup | |
| Call Supplier |
Below Normal | YELLOW |
Normal O2 | GREEN |
FIGURE 7 - O2 DISPLAY CONTROL PANEL