HoMedics BPA-100 manual Potential for RADIO/TELEVISION Interference, Troubleshooting

Page 15

BPA-100IB_ENGSP.qxd 10/24/05 10:51 AM Page 27

POTENTIAL FOR RADIO/TELEVISION

INTERFERENCE

For USA only:

This product has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. The product generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If the product does cause harmful interference to radio or television reception, which can be determined by turning the product on and off, the user is encouraged to try to correct the interference by one or more of the following measures:

Reorient or relocate the receiving antenna.

Increase the separation between the product and the receiver.

Connect the product into an outlet on a circuit different from that to which the receiver is connected.

Consult the dealer or an experienced radio/TV technician for help.

For Canada only:

This digital apparatus does not exceed the

Class B limits for radio noise emissions from digital apparatus as set out in the interference-causing equipment standard entitled “Digital Apparatus”, ICES-003 of the Canadian Department of Communications.

Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux appareils numériques de Classe B prescrites dans la norme sur le matériel brouilleur: “Appareils Numériques”, ICES-003 édictée par le ministre des communications.

TROUBLESHOOTING

If any abnormality arises during use, please check the following points

SYMPTOMS

POSSIBLE

CORRECTION

CAUSES

 

 

 

 

 

Unit does not turn on

Batteries have

Replace them with

when the START/STOP

run down.

four new AA

button is pushed.

 

alkaline batteries.

 

 

 

 

Battery polarities

Re-insert the batteries in

 

have been positioned

the correct positions.

 

incorrectly.

 

 

 

 

EE measurement error

Cuff has been placed

Wrap the cuff properly

symbol shown on dis-

incorrectly.

so that it is positioned

play or the blood pres-

 

correctly.

sure value is displayed

 

 

Did you talk or move

Measure again. Keep

excessively low

during measurement?

arm steady during

(or high).

 

measurement.

Shaking of the arm with

 

 

 

the cuff on.

 

 

 

 

E1 error symbol shown

Air circuit abnormality.

Check cuff connection.

on display

Cuff tube may not be

Measure again.

 

plugged into monitor

 

 

correctly.

 

 

 

 

E2 error symbol shown

Inflation pressure

Switch the unit off, then

on display

exceeding 300 mmHg.

measure again.

 

 

 

E3 error symbol shown

Data error.

Remove and reload the

on display

 

batteries.

 

 

 

Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance should you disassemble or attempt to repair the unit by yourself. Contact information for HoMedics Consumer Relations Department can be found on the warranty page.

27

28

Image 15
Contents Automatic Blood Pressure Monitor Table of Contents Important Product Notices and Safety Instructions About Blood Pressure NAME/FUNCTION of Each Part Monitor WorksDisplay Explanations Display SymbolsUsing AN AC Adapter not Included Installing BatteriesApplying the Cuff Date & Time SET ProcedurePull end toward you, then wrap over cuff Important Notes Measurement ProcedurePress the User Select Button to choose User 1 or User Recalling Values from Memory WHO Health IndicatorImportant Notes Regarding Your Blood Pressure Measurement Clearing Values from MemoryCARE, Maintenance & Cleaning Potential for RADIO/TELEVISION Interference TroubleshootingDimensions SpecificationsMonitor automático de presión sanguínea Información DE USO Información InicialAvisos Importantes DEL Producto E Instrucciones DE Seguridad Sobre LA Presión Sanguínea NOMBRE/FUNCIÓN DE Cada Pieza Símbolos de la pantalla Explicaciones DE LA PantallaUSO DE UN Adaptador DE CA no Incluido Instalación DE LAS PilasAplicación DEL Brazalete Procedimiento DE Configuración DE Fecha Y HoraMain Arteria Artery principal Notas importantes Procedimiento DE MediciónAutomáticamente en la memoria Indicador DE Salud DE LA OMS Cómo Borrar LOS Valores DE LA Memoria CUIDADO, Mantenimiento Y Limpieza Potencial DE Interferencia CON RADIO/TELEVISIÓN Diagnóstico Y Resolución DE ProblemasDesinflado Especificaciones