Invacare RPS350-2 manual Replacing Casters/Forks, Replacing Rear Casters

Page 36

MAINTENANCE

Replacing Casters/Forks

Replacing Rear Casters

NOTE: For this procedure, refer to FIGURE 7.7

1.Place the lift on its side.

2.Remove the bolt and locknut that secure the existing rear caster to the fork.

NOTE: The bushing will be loose and may fall out of the caster.

NOTE: Existing bushing will be reused. Examine and replace if worn.

3.Install the new/existing bushing into the new rear caster.

4.Line up the mounting holes in the new rear caster and the fork.

5.Install the bolt through the fork and new rear caster and tighten securely with the locknut.

Base

Fork

Locknut

Rear Caster

Bolt

Bushing

FIGURE 7.7 Replacing Rear Casters

Replacing Front Casters

NOTE: For this procedure, refer to FIGURE 7.8.

1.Place the lift on its side.

2.Remove the bolt and locknut that secure the existing front caster assembly to the fork.

NOTE: The front caster assembly consists of two casters and a washer in between.

NOTE: The washer will fall out from between

Base

the two casters.

 

NOTE: Washer will be reused. Examine and

replace if worn.

Fork

3.

Position the new/existing washer

Locknut

 

between the two new casters.

 

4.

Line up the mounting holes in the new

 

 

front caster assembly and the fork.

Bolt

5.

Install the bolt through the fork and the

Washer

 

new front caster assembly and tighten

 

securely with the locknut.

Front Casters

Image 36
Contents Stand Up Lift Model RPS350-2 Stand Up Lift Model RPS350-2 Table of Contents Maintenance Register Your ProductProduct Registration Form Fold here Cut Along Line Special Notes MeaningSpecifications Patient LiftGeneral Guidelines Assembling the LiftOperating the Lift Using the Sling Lifting/Transferring the Patient Performing Maintenance Electrical Grounding InstructionsPinch Points Installation IntroductionAssembling the Mast to the Base Installing the Leg Actuator to the Base Assembling the Mast ActuatorLift Components Mounting the Battery Charger Mounting the Battery ChargerUsing the Pendant Buttons Operating the Stand UP LiftPerforming an Emergency Stop Raising/Lowering the Stand Up LiftCharging the Battery Charging the BatteryLifting the Patient Positioning the Stand Up LiftLifting the Patient Lifting the Patient Moving the Patient Lifting the PatientTransferring the Patient Transferring to a Commode Transferring to a CommodeTransferring to a Wheelchair Transferring to a Bed Transferring the Patient to a WheelchairTransferring the Patient to a Bed Troubleshooting Symptoms Faults SolutionMaintenance Maintenance Safety Inspection ChecklistCleaning the Sling and the Lift Lubricating the LiftDetecting Wear and Damage Adjusting the Base Adjusting the Knee Pad HeightReplacing the Knee Pad Replacing the Padded CoverReplacing the Mast Actuator Replacing the Mast ActuatorReplacing Casters/Forks Replacing Rear CastersReplacing Forks Maintenance Limited Warranty USA