Invacare Spectra Plus Controller Options, DUAL/ATTENDANT Control, Circuit Breakers 30 AMP

Page 19

CONTROLLER OPTIONS

The optional Attendant, Chin and Centre are fitted with the PILOT+ Control System. The display function is the same for each joystick module.

DUAL/ATTENDANT CONTROL

CAUTION: ATTENDANT, Before switching from Occupant Control to Attendant control, it is of the utmost importance that the speed control is checked and a safe manageable speed is selected before activating the joystick.

Failure to do so may cause the chair to move uncontrollably, causing injury to the occupant or other people.

CHIN CONTROL

ATTENDANT CONTROL

CIRCUIT BREAKERS - 30 AMP

The circuit breakers are located at the front of the front battery box and at the front of the rear battery box between the two battery boxes. They are fitted to operate if a current overload should occur. If one of the breakers operate, it will ‘pop’ out. To re-set, push the knob in after a few minutes. If the circuit breaker continually fails, consult your authorised INVACARE dealer.

Under certain circumstances, particularly indoor use, care must be taken to ensure safe continual operation of the motor units. If an obstacle is encountered or, with slow manoeuvring of the castor wheels from the forward to rear position, continual full operation of the joystick control will result in motor overload and eventual failure.

To prevent this occurrence, increase the speed to maximum, control the manoeuvre by the joystick control, always move away from the obstacle, do not continue to drive against it.

17

Image 19
Contents Spectra Plus Contents Advisory Safety Notice Purpose of the Power ChairForeword Safety Precautions Getting Acquainted with Your NEW Power ChairJoystick FeaturesSeat Depth Adjustment 35mm depth adjustment Seat and Backrest AdjustmentBackrest Angle Adjustment Armrest Adjustment MANUALALTILTT--IN--SPACE TensiontensionadjustableadjustablebackrestPilot Integral Handling and Operation ControllersRecommended Driving Position Dynamic DL ControllerDriving the Wheelchair Important Information for Your Safety IntroductionDX-REMG80IMASTER Remote Handling and TransportationFault Diagnostics MaintenanceSafety and Misuse Warnings Side Lights Controlling Lighting Functions Using the KeypadLeft and Right Indicators Appendix aNegotiating GRADIENTS/SLOPES PILOT+Negotiating Kerbs Controls Joystick Module DetailsHorn Switch Maximum Speed Indicator Speed Adjustment Mode Chin Control Attendant Control Controller OptionsDUAL/ATTENDANT Control Circuit Breakers 30 AMPTransfers For Clearance Elevating Legrests AdjustmentFootrests Lowest Part Footrest Must be AT LeastLevers Pointing Rearwards Pushing the Chair Freewheel ModeGoing Down a Kerb RE-ASSEMBLYOperating the Chair Outdoors Kerb Climbing Going UP a KerbDIS-ASSEMBLY Detachable Motors Drive WHEEL-LOCKS Manual Brakes Drive Wheel RemovalLabelled „TIIEDOWNPOINTPOINT Diiagram Fig. No..2 shows Power chair was ttestted toOccupant’s inertia seat belt Which also must be used Fig. No.3 TransportationOperation Removal Switch OFF the Controller Tray Assembly Fitting Operation InstructionsShown in Fig Lateral Support Fitting InstructionsBatteries Charging BatteriesMains Socket Charger Controller Front Pollution from improper disposal Battery ConnectionsCHROME/PAINTWORK Keeping Your Chair in Good ConditionFault Finding Technical Information Spectra PlusDaily Weekly Monthly Yearly Approved Accessories Available Models Covered Spectra Plus Spectra Plus