SECTION 11—BATTERIES
NOTE: For this procedure, refer to FIGURE 11.7 on page 54.
1.Install battery terminal cap(s) onto battery cable(s):
A.ORANGE battery terminal cap onto RED battery cable.
B.GREY battery terminal cap onto BLACK battery cable.
CAUTION
When connecting the battery cables to the battery(ies), the battery cable(s) MUST be connected to the battery terminal(s)/post(s) as shown in Detail “A” or Detail “B” of FIGURE 11.7 (depending on battery type). Otherwise, damage to the battery cable may result when installing battery terminal caps.
2.Connect battery cable(s) to battery(ies) terminal(s)/post(s) as shown in Detail “A”:
A.NEGATIVE
B.POSITIVE (+) RED battery cable to POSITIVE (+) battery terminal/post.
3.Secure the battery cable(s)/ring terminal(s) to the battery terminal(s)/post(s), BLACK to NEGATIVE
4.Verify all battery cable(s)/ring terminal(s) are correctly installed and securely tightened.
5.Slide terminal cap(s) down battery cable(s) and onto battery clamps.
6.Using the
NOTE: It will be necessary to trim excess
7.Install the battery box top(s).
8.Install the battery box(es) into the wheelchair. Refer to Removing/Installing the Battery Boxes on page 49.
NOTE: New battery(ies) MUST be fully charged before using, otherwise the life of the battery(ies) will be reduced.
9. If necessary, charge the battery(ies). Refer to Charging Batteries on page 47.
Part No. 1123839 | 53 |