ResMed VPAP III ST-A Helpful Hints, Servicing, Travelling with the Vpap III ST-A with QuickNav

Page 17

CAUTION

Do not attempt to open the device. There are no user serviceable parts inside. Repairs and internal servicing should only be performed by an authorised service agent.

Replacing the Air Filter

Inspect the air filter every month to check if it is blocked by dirt or contains holes. With normal use of the device, the air filter needs to be replaced every six months (or more often if your device is in a dusty environment). To replace the air filter:

1Remove the air filter cover at the back of the device.

2Remove and discard the old air filter.

3Insert a new filter with the blue tinted side facing out.

4Replace the air filter cover.

WARNING

Do not wash the air filter. The air filter is not washable or reusable.

Servicing

This product (VPAP III ST-A with QuickNav) should be inspected by an authorised ResMed service centre five years from the date of manufacture. Prior to this, the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed. Applicable ResMed warranty details are provided with the device at the time of original supply. Of course, as with all electrical devices, if any irregularity becomes apparent, you should exercise caution and have the device inspected by an authorised ResMed service centre.

If you feel that your device is not performing properly, see “Troubleshooting” on page 15.

CAUTION

!Inspection and repair should only be performed by an authorised agent. Under no circumstances should you attempt to service or repair the flow generator yourself.

Helpful Hints

Travelling with the VPAP III ST-A with QuickNav

International Use

Your device has an internal power adapter that enables it to operate in other countries. It will operate on power supplies of 100–240V, 50–60Hz and 110V, 400Hz. No special adjustment is necessary, but you will require an approved power cord for that country.

Using a Battery to Power the Device

Information regarding suitable DC and battery power supplies for this device can be found

on www.resmed.com on the Products page under Service and Support.

14

Image 17
Contents Vpap III ST-A with QuickNav Vpap III ST-A with QuickNav Guidance and Manufacturer’s Declaration Electromagnetic Medical Information IntroductionMasks Vpap III ST-A with QuickNavHumidifiers SupportSetting up the device Attaching a Humidifier Using the LCD QuickNav Screen and KeypadHumidAire 2i/2iC Humidifier HumidAire or Passover HumidifierKeypad Keys Starting TreatmentLEDs Patient Standby Screens Using the HumidAire 2i Warm-up FeatureRamp Time Patient Home Screen/Treatment ScreenStopping Treatment Using the Mask-Fit FeatureQuickNav Screen How to Use the MenusMenu Navigation Patient Treatment MenuFunction Default Function Description Settings Patient Settings MenuPatient Options Menu Function Function Description Default SettingsAlarms Troubleshooting AlarmsAlarm Mute Key Testing the AlarmLCD Ipap Lower Daily Cleaning Mask Cleaning and MaintenanceWeekly Cleaning Periodic CleaningHelpful Hints Using a Battery to Power the DeviceServicing Travelling with the Vpap III ST-A with QuickNavTroubleshooting System Specifications Dimensions L x W x H Sound power levelWeight Air outlet Power supplySymbols which may Appear on the Product General Warnings and Cautions Page Separation distance according to frequency of transmitter m Product Warranty Period Limited WarrantyManufacturer Distributed by Eur 1 ENG