Problem / Possible Cause | Solution | |
|
|
|
| Breath is not deep enough to trigger | Take a deep breath in and out through the |
| SmartStart. | mask. |
|
|
|
| There is excessive leak. | Adjust position of mask and headgear. |
|
|
|
| Plugs may be missing from ports on mask. | Replace plug(s). |
|
|
|
| Air tubing is not connected properly. | Connect firmly at both ends. |
|
|
|
| Air tubing is kinked or punctured. | Straighten or replace. |
|
|
|
| There is a large impedance (eg, antibacterial | Press the Start/Stop key. |
| filter, oxygen connector) in the air circuit. |
|
The device does not stop when you remove your mask.
| SmartStart/Stop is disabled. | Enable SmartStart/Stop. |
| Use of a full face mask. | SmartStop does not work with a full face |
|
| mask. |
|
|
|
| Incompatible humidifier or mask system | Use only equipment as recommended and |
| being used. | supplied by ResMed. |
|
|
|
| Leak Alarm or Low MV Alarm is set to ON. | Consult your clinician. |
| ||
Displays error message: SYSTEM ERROR Call service! | ||
|
|
|
| Component failure. | Return your device for servicing. |
|
| |
Excessive motor noise. |
| |
|
|
|
| Component failure. | Return your device for servicing. |
|
|
|
System Specifications
Dynamic pressure | • | IPAP: 2 cm H2O to 30 cm H2O (measured at the end of | |
characteristics: |
| standard 2 m air tubing) |
|
|
| ||
| • EPAP: 2 cm H2O to 25 cm H2O (measured at the end of | ||
|
| standard 2 m air tubing) |
|
|
| ||
| • CPAP: 4 cm H2O to 20 cm H2O (measured at the end of | ||
|
| standard 2 m air tubing) |
|
|
|
| |
Maximum single fault | 40 cm H2O |
| |
pressure: |
|
|
|
|
|
| |
Maximum flow | Pressure (cm H2O) | Flow (L/min) | |
(Pressure, measured at |
|
|
|
4 |
| 244 | |
the end of standard |
| ||
|
|
| |
2 m air tubing) | 8 |
| 252 |
|
|
|
|
| 12 |
| 245 |
|
|
|
|
| 16 |
| 248 |
|
|
|
|
| 20 |
| 246 |
|
| ||
Sound pressure level: | <30 dB (tested in accordance with the requirements of ISO | ||
|
|
<37 dB (tested in accordance with the requirements of ISO
16