ResMed VPAP III ST-A manual Starting Treatment, LEDs, Keypad Keys

Page 8

To assist you in adjusting the device, the keypad and LCD are equipped with a backlight. The LCD backlight comes on when the device is turned on or when you press a key, and turns off after two minutes. The LCD backlight may be set to be on continually. The keypad backlight is on at all times when the device is powered.

LEDs

The Therapy LED (white) may be on during treatment, if enabled by your clinician. The Alarm LEDs (Red/Yellow) are on during an alarm condition or during alarm testing.

Keypad Keys

The device keypad has the following keys:

Key

Function

 

 

 

Start/Stop

• Starts or stops treatment.

 

 

• Extended hold for at least two seconds starts the Mask­Fitting

 

 

 

 

feature.

 

 

 

QuickView

• Clinical menu function key. Clinical use only.

 

 

 

Alarm Mute

• Press once to mute alarms. Press a second time to un-mute

 

 

alarms. If the problem is still present, the alarm will sound again

 

 

after two minutes. See “Alarm Mute Key” on page 11.

 

 

 

Exit (red)

• Allows you to exit the current menu or go back through the

 

 

menus. The function of this key is to exit from the current menu

 

 

or setting.

Up/Down

• Allows you to scroll through the device menus, submenus and

 

 

settings.

English

Enter (green)

• Allows you to enter or change the menu or function highlighted

 

on the LCD screen. Functions of this key includes enter, change

 

and apply, and it also operates as a soft key.

 

 

Starting Treatment

The device should be assembled beside your bed with the air tubing and mask system connected.

1Turn the power switch at the back of the device to on (I).

When the device is turned on, the Patient Standby screen is displayed.

Notes:

If you have the HumidAire 2i attached, see “Using the HumidAire 2i Warm-up Feature” on page 6.

When the device is turned on, the alarm will sound a test beep and the red and yellow LEDs will flash. To test the alarm manually, or to change the volume, see “Testing the Alarm” on page 11.

2Fit your mask as described in the mask user instructions.

Starting Treatment

5

Image 8
Contents Vpap III ST-A with QuickNav Vpap III ST-A with QuickNav Guidance and Manufacturer’s Declaration Electromagnetic Introduction Medical InformationVpap III ST-A with QuickNav MasksHumidifiers SupportSetting up the device Using the LCD QuickNav Screen and Keypad Attaching a HumidifierHumidAire 2i/2iC Humidifier HumidAire or Passover HumidifierKeypad Keys Starting TreatmentLEDs Using the HumidAire 2i Warm-up Feature Patient Standby ScreensRamp Time Patient Home Screen/Treatment ScreenUsing the Mask-Fit Feature Stopping TreatmentHow to Use the Menus QuickNav ScreenMenu Navigation Patient Treatment MenuPatient Settings Menu Function Default Function Description SettingsFunction Function Description Default Settings Patient Options MenuAlarms Alarms TroubleshootingAlarm Mute Key Testing the AlarmLCD Ipap Lower Cleaning and Maintenance Daily Cleaning MaskWeekly Cleaning Periodic CleaningUsing a Battery to Power the Device Helpful HintsServicing Travelling with the Vpap III ST-A with QuickNavTroubleshooting System Specifications Sound power level Dimensions L x W x HWeight Air outlet Power supplySymbols which may Appear on the Product General Warnings and Cautions Page Separation distance according to frequency of transmitter m Limited Warranty Product Warranty PeriodManufacturer Distributed by Eur 1 ENG