ResMed VPAP III ST-A manual Alarms Troubleshooting, Alarm Mute Key, Testing the Alarm

Page 14

The Alarms

This device is fitted with alarms to alert you to changes that will affect your treatment.

Alarm Mute Key

You can mute an alarm by pressing the Alarm Mute key once. Pressing this key a second time will un-mute the alarm. If the problem is still present, the alarm will sound again after two minutes. An alarm LED will remain lit for as long as the problem is present. The Title bar of the LCD will display the alarm description while the alarm is muted.

Testing the Alarm

When the device is turned on, the LEDs will flash and the alarm will beep to confirm that the alarm is working.

Setting the Alarm Volume

You can set and test the alarm volume from the Patient Options Menu. The alarm will beep at the selected volume when you press Enter.

All menus remain available during an alarm. Treatment screens are not viewable during an alarm condition.

Alarms Troubleshooting

The most common reason for an alarm to sound is because the system has not been properly assembled. Check that the air tubing has been properly attached to the flow generator and mask (and humidifier if used).

When an alarm has been activated, the LCD screen will display either alarm information or recommended instructions for you or your clinician.

Note: The alarm actions listed below are based on having the appropriate alarm settings for your therapy. When an alarm is activated, consult your clinician.

CAUTION

!In the event of power failure or machine malfunction, remove the mask to avoid rebreathing your exhaled air.

Warning Signal/Cause

Action

 

 

For all the medium priority alarms listed below, you will hear a single intermittent tone and

the yellow LED will flash.

LCD: LCD turns off

The flow generator stops delivering air pressure.

• Power failure.

Remove your mask until power is restored.

• Power cord is disconnected or

Notes:

device switched off while

Treatment will re-start when power is restored.

delivering treatment (without

Unless muted, the alarm will sound for at least

two minutes in the event of a power failure.

pressing the Start/Stop key).

 

LCD: CHECK TUBE

The flow generator stops delivering air pressure.

• Air tubing disconnected from the

1.

Check that the air tubing is connected

HumidAire 2i/2iC.

 

properly to the HumidAire 2i/2iC.

 

2.

Check that the HumidAire 2i/2iC or front cover

 

 

is connected properly to the flow generator.

English

The Alarms

11

Image 14
Contents Vpap III ST-A with QuickNav Vpap III ST-A with QuickNav Guidance and Manufacturer’s Declaration Electromagnetic Introduction Medical InformationHumidifiers Vpap III ST-A with QuickNavMasks SupportSetting up the device HumidAire 2i/2iC Humidifier Using the LCD QuickNav Screen and KeypadAttaching a Humidifier HumidAire or Passover HumidifierKeypad Keys Starting TreatmentLEDs Ramp Time Using the HumidAire 2i Warm-up FeaturePatient Standby Screens Patient Home Screen/Treatment ScreenUsing the Mask-Fit Feature Stopping TreatmentMenu Navigation How to Use the MenusQuickNav Screen Patient Treatment MenuPatient Settings Menu Function Default Function Description SettingsFunction Function Description Default Settings Patient Options MenuAlarm Mute Key AlarmsAlarms Troubleshooting Testing the AlarmLCD Ipap Lower Weekly Cleaning Cleaning and MaintenanceDaily Cleaning Mask Periodic CleaningServicing Using a Battery to Power the DeviceHelpful Hints Travelling with the Vpap III ST-A with QuickNavTroubleshooting System Specifications Weight Air outlet Sound power levelDimensions L x W x H Power supplySymbols which may Appear on the Product General Warnings and Cautions Page Separation distance according to frequency of transmitter m Limited Warranty Product Warranty PeriodManufacturer Distributed by Eur 1 ENG
Related manuals
Manual 56 pages 60.78 Kb