ResMed VPAP III ST-A manual Troubleshooting

Page 18

Inverter or UPS

The power specifications for an inverter or a UPS are listed in the following table.

Configuration

Continuous Output

Peak Output

Power Rating

Power Rating

 

 

 

 

 

VPAP III ST-A with QuickNav (without

60W

225W

 

HumidAire 2i humidifier)

 

 

 

 

 

 

 

VPAP III ST-A with QuickNav with

150W

300W

(110V inverter)

HumidAire 2i humidifier

 

600W

(240V inverter)

Note: Use only a pure sine wave

 

 

 

inverter when a HumidAire 2i is attached.

The case temperature should be less than 50°C at an ambient temperature of 35°C. (For VPAP III ST-A with QuickNav temperature specifications, see “System Specifications” on page 16.)

Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device.

Problem / Possible Cause

Solution

 

 

 

No display.

 

 

 

 

 

Power not connected or switch at back is

Ensure the power cable is connected and

 

not on.

that the switch at the back of the device is

 

 

in the ON position.

 

 

Insufficient air delivered from the device.

 

 

 

 

Ramp Time is in use.

Wait for air pressure to build up.

 

 

 

 

Air filter is dirty.

Replace air filter.

 

 

 

 

Air tubing is kinked or punctured.

Straighten or replace tubing.

 

 

 

 

Air tubing is not connected properly.

Check air tubing.

 

 

 

 

Mask and headgear not positioned

Adjust position of mask and headgear.

 

correctly.

 

 

 

 

 

Plug(s) missing from access port(s) on

Replace plug(s).

 

mask.

 

 

 

 

 

Pressure required for treatment may have

See your clinician to adjust the pressure.

 

changed.

 

 

The device does not start when you breathe into the mask (when

SmartStart is enabled).

 

 

 

 

 

Power cord not connected properly.

Connect power cord firmly at both ends.

 

 

 

 

Power outlet may be faulty.

Try another power outlet.

 

 

 

 

The device is not switched on.

Turn power switch at rear of the device to

 

 

on (I).

 

 

 

 

SmartStart not on.

Enable SmartStart.

 

 

 

English

Troubleshooting 15

Image 18
Contents Vpap III ST-A with QuickNav Vpap III ST-A with QuickNav Guidance and Manufacturer’s Declaration Electromagnetic Introduction Medical InformationHumidifiers Vpap III ST-A with QuickNavMasks SupportSetting up the device HumidAire 2i/2iC Humidifier Using the LCD QuickNav Screen and KeypadAttaching a Humidifier HumidAire or Passover HumidifierStarting Treatment LEDsKeypad Keys Ramp Time Using the HumidAire 2i Warm-up FeaturePatient Standby Screens Patient Home Screen/Treatment ScreenUsing the Mask-Fit Feature Stopping TreatmentMenu Navigation How to Use the MenusQuickNav Screen Patient Treatment MenuPatient Settings Menu Function Default Function Description SettingsFunction Function Description Default Settings Patient Options MenuAlarm Mute Key AlarmsAlarms Troubleshooting Testing the AlarmLCD Ipap Lower Weekly Cleaning Cleaning and MaintenanceDaily Cleaning Mask Periodic CleaningServicing Using a Battery to Power the DeviceHelpful Hints Travelling with the Vpap III ST-A with QuickNavTroubleshooting System Specifications Weight Air outlet Sound power levelDimensions L x W x H Power supplySymbols which may Appear on the Product General Warnings and Cautions Page Separation distance according to frequency of transmitter m Limited Warranty Product Warranty PeriodManufacturer Distributed by Eur 1 ENG
Related manuals
Manual 56 pages 60.78 Kb