ResMed VPAP III ST-A Sound power level, Dimensions L x W x H, Weight Air outlet, Flow measurement

Page 20

Sound power level:

<45 dB (tested in accordance with the requirements of ISO

 

17510-1:2007)

 

 

Dimensions (L x W x H):

270 mm x 230 mm x 141 mm

 

 

Weight:

2.3 kg

 

 

Air outlet:

22 mm taper, compatible with ISO 5356-1:2004 Anaesthetic &

 

Respiratory Equipment - Conical Connectors

 

 

Pressure measurement:

Internally mounted pressure transducer

 

 

Flow measurement:

Internally mounted flow transducer

 

 

Power supply:

AC 100–240V, 50–60Hz, 2.2A; AC 110V, 400 Hz, 2.2A;

 

DC 24V, 2A

 

 

Housing construction:

Flame retardant engineering thermoplastic

 

 

Environmental

• Operating Temperature: +5°C to +35°C

conditions:

 

 

• Operating Humidity: 10%–95% non-condensing

 

 

 

 

• Storage and Transport Temperature: -20°C to +60°C

 

 

 

• Storage and Transport Humidity: 10%–95% non-condensing

 

 

Electromagnetic

Product complies with all applicable electromagnetic

compatibility:

compatibility requirements (EMC) according to IEC60601-1-2,

 

for residential, commercial and light industry environments. For

 

further details, see “Guidance and Manufacturer’s Declaration -

 

Electromagnetic Emissions and Immunity” on page 20.

 

 

Air filter:

Two-layered, powder-bonded, polyester non-woven fibre

 

 

Air tubing:

Flexible plastic, 2 m or 3 m length

 

 

IEC 60601-1

• Class II (double insulation)

classifications:

 

 

Type CF

 

 

 

 

 

Continuous operation

This flow generator is not suitable for use in the presence of a flammable anaesthetic mixture with air, or with oxygen or nitrous oxide entrained in the flow generator airpath.

Displayed values

Value

Range

Accuracy

Display Resolution

 

 

 

Pressure sensor at air outlet

 

 

 

 

 

 

Pressure

-5 to 30 cm H2O

±0.5 cm H2O (+ 4% of

0.1 cm H2O

 

 

measured value)

 

 

 

 

Flow sensor in flow generator*

 

 

 

 

 

 

Leak

0–120 L/min

**

1 L/min

 

 

 

 

Tidal volume

50–3,000 mL

**

1 mL

 

 

 

 

Respiratory rate

6–60 BPM

±0.5 BPM

1 BPM

Minute ventilation

0.6–60 L/min

**

0.1 L/min

 

 

 

 

*Results may be inaccurate in the presence of leaks or supplemental oxygen.

**The displayed values are estimates. They are provided for trending purposes only.

† Results may be inaccurate if the tidal volume is below 50mL.

English

System Specifications

17

Image 20
Contents Vpap III ST-A with QuickNav Vpap III ST-A with QuickNav Guidance and Manufacturer’s Declaration Electromagnetic Introduction Medical InformationVpap III ST-A with QuickNav MasksHumidifiers SupportSetting up the device Using the LCD QuickNav Screen and Keypad Attaching a HumidifierHumidAire 2i/2iC Humidifier HumidAire or Passover HumidifierKeypad Keys Starting TreatmentLEDs Using the HumidAire 2i Warm-up Feature Patient Standby ScreensRamp Time Patient Home Screen/Treatment ScreenUsing the Mask-Fit Feature Stopping TreatmentHow to Use the Menus QuickNav ScreenMenu Navigation Patient Treatment MenuPatient Settings Menu Function Default Function Description SettingsFunction Function Description Default Settings Patient Options MenuAlarms Alarms TroubleshootingAlarm Mute Key Testing the AlarmLCD Ipap Lower Cleaning and Maintenance Daily Cleaning MaskWeekly Cleaning Periodic Cleaning Using a Battery to Power the Device Helpful Hints Servicing Travelling with the Vpap III ST-A with QuickNavTroubleshooting System Specifications Sound power level Dimensions L x W x HWeight Air outlet Power supplySymbols which may Appear on the Product General Warnings and Cautions Page Separation distance according to frequency of transmitter m Limited Warranty Product Warranty PeriodManufacturer Distributed by Eur 1 ENG