SECTION 3—CHECKING O2 PURITY

SECTION 3—CHECKING O2 PURITY

NOTE: Although the XPO2 offers pulse dose only, the oxygen purity can be checked with a hand held oxygen analyzer.

NOTE: Oxygen purity should be checked every 4,320 hours.

1.Turn the unit On.

2.Press the “+” button to increase the flow to setting 5.

3.Simultaneously press and hold the “+” and “battery gauge” buttons until the #5 BLUE indicator light blinks. This is your indication that the unit is now in auto-pulse mode.

NOTE: The unit will remain in auto-pulse mode until it is set to a lower flow setting or is turned off.

4.Connect a standard hand held oxygen analyzer to the outlet port.

5.Follow the directions provided by the analyzer manufacturer. Some analyzers will also measure flow and pressure, but only if connected to a continuous flow of oxygen (so these options are not usable on this pulse only unit).

NOTE: The purity reading may fluctuate slightly due to the pulse flow. It is recommended that you record a high and a low over a 60 second time period. The XPO2 is functioning properly when the purity levels are above 87%, after the 5 minute warm-up period.

XPO2

14

Part No 1150748

Image 14
Contents XPO2 Accessories Warning Table of Contents Special Notes Typical Product Parameters IPX1Regulatory Listing Safeguards Operating InformationRadio Frequency Interference Component Replacement Accessing XPO2Removing the Front Cover Replacing the Inlet Filter Replacing the Inlet FilterReplacing the Battery Replacing the BatteryReplacement Parts Re-assembling the XPO2Checking O2 Purity Error Codes Common AlarmsCompressor Alarms XPO2 XPO2 Limited Warranty USA