Shop-Vac XPO100B service manual Replacing the Inlet Filter

Page 11

SECTION 2—COMPONENT REPLACEMENT

4.While holding the unit together rotate the XPO2 and position it so that the back panel label is facing down.

NOTE: The ribbon cable is connected to the front cover and the pc board. Only open the XPO2 as shown in Details “B” and “C”.

5.Slowly, open the front cover ensuring not to damage the ribbon cable (Detail “B” and Detail “C” in FIGURE 2.3).

Replacing the Inlet Filter

NOTE: For this procedure, refer to FIGURE 2.4 and FIGURE 2.2 on page 10.

NOTE: The inlet filter should be replaced as needed.

1.Perform the steps in Accessing XPO2 on page 9.

2.Remove the inlet filter from the holder (Detail “A”).

3.Disconnect the inlet filter with 5/8-inch tubing attached from the 111/2-inch long tubing (Detail “B”).

Holder

DETAIL “A”

Inlet Filter

DETAIL “B”

5/8-Inch Tubing

Intake

Inlet Filter w/ 5/8-inch

 

Attached

 

Flow Arrow

 

Outlet

111/2-inch Tubing

FIGURE 2.4 Replacing the Inlet Filter

Part No 1150748

11

XPO2

Image 11
Contents XPO2 Accessories Warning Table of Contents Special Notes IPX1 Typical Product ParametersRegulatory Listing Operating Information SafeguardsRadio Frequency Interference Accessing XPO2 Component ReplacementRemoving the Front Cover Replacing the Inlet Filter Replacing the Inlet FilterReplacing the Battery Replacing the BatteryRe-assembling the XPO2 Replacement PartsChecking O2 Purity Common Alarms Error CodesCompressor Alarms XPO2 XPO2 Limited Warranty USA