SECTION
NOTE: When installing new battery, ensure the battery is label side up.
5.Position new battery on thermal pad as shown.
6.Connect the battery wiring harness to the PC board.
7.Install inlet filter with tubing into holder.
8.
Re-assembling the XPO2
NOTE: For this procedure, refer to FIGURE 2.3 on page 10.
1.Carefully,
2.Turn the XPO2 On to ensure proper operation.
3.Turn XPO2 Off.
4.While holding the unit together, position the XPO2 so that the back panel is facing up (Detail “A” in FIGURE 2.3).
5.Install four mounting screws and secure the two sections of the XPO2 together (Detail “A” in FIGURE 2.3). Securely tighten.
6.Remove the glue residue by applying 91% isopropyl alcohol, and wiping with a clean cloth. Make sure the surface is dry and free of any residue. This will ensure a good seal for the new label.
7.Install replacement label on the top of the XPO2 so that it adheres to both the front and rear sections.
Replacement Parts
NOTE: Refer to the parts catalog at www.invacare.com for component part number.
Part No 1150748 | 13 | XPO2™ |