Invacare XPO100B user manual Battery is charging while unit is Running

Page 15

SECTION 5—TYPICAL PRODUCT PARAMETERS

Oxygen Concentration:*

87% to 95.6%, after initial warm-up

*Based on an atmospheric

period (approximately 5 minutes)

pressure of 14.7 psi (101 kPa) at

 

70°F (21°C)

 

 

 

Oxygen Flow:

Pulse flow delivery. Bolus Volume Ranges

 

From 300 - 840cc settings 1 to 5.

 

 

Dimensions:

10” high x 7” wide x 4” deep (25.4 cm

 

high x 17.8 cm wide x 10.2 cm wide)

 

 

Weight:

6.0 lbs (2.72 kg); 1.3 lbs (0.68 kg) for

 

supplemental battery

 

 

Battery Duration (times are

Setting 1 = 3.5 hrs

approximate):

 

Setting 2 = 2.5 hrs

 

 

 

 

Setting 3 = 2.0 hrs

 

 

 

Setting 4 = 1.5 hrs

 

 

 

Setting 5 = 1.0 hrs

 

 

Battery Recharge Time:

4 hours

NOTE: Recharge time increases if

 

battery is charging while unit is

 

running.

 

 

 

Humidity

Operating Humidity: 15% to 60% non

 

condensing

 

 

 

Storage Humidity: up to 95% non

 

condensing

 

 

Temperature Range:

Operating temperature: 41°F to 95°F

(All power sources)

(5°C to 35°C)

 

 

 

Storage temperature: -2°F to 140°F

 

(-20°C to 60°C)

 

 

Extended Temperature Range:

Operating Temperature:

(Using AC or DC adapters)

95°F to 104° (35°C to 40°C)

 

Continuous use - all settings

 

 

Extended Temperature Range

Operating Temperature:

(Using Internal Battery)

95°F to 104° (35°C to 40°C)

 

Continuous use - settings 1, 2 and 3

 

45 minutes (max) - setting 4

 

30 minutes (max) - setting 5

 

 

Part No 1148112

15

Invacare XPO2

Image 15
Contents Invacare XPO2 Portable Concentrator Accessories Warning Table of Contents Table of Contents Special Notes Disposal of Equipment and Accessories General Guidelines Operating Information Radio Frequency Interference MaintenancePolarized Plug Instruction Storage Packaging and HandlingUnpacking InspectionTechnical Description XPO2 Features XPO2FEATURESIPX1 Typical Product ParametersBattery is charging while unit is Running Regulatory Listing Filters IntroductionRecommendations for Optimal Performance LocationPower Supplies Power Supplies Checking Charge Level of Internal Battery Checking Charge Level of Internal BatteryCharging the Batteries Charging the Internal BatteryCharging the Supplemental Battery Only Charging the Supplemental Battery Only Operating Instructions Prolonging XPO2 Battery Life Battery Time ManagementDon’ts Powering the XPO2 Powering the XPO2 Connecting/Positioning the Nasal Cannula Connecting/Positioning the Nasal Cannula Adjusting the Pulse Flow SettingAdjusting the Pulse Flow Setting Using the XPO2 Portable ConcentratorIndicators, Warnings and Alarms Operating Instructions Operating Instructions Operating Instructions Cabinet CLEANING, CARE, and MaintenanceAir Intake Filter Air Intake FilterCarrying Bag Carrying BagALARM/WARNING Symptom Probable Cause Solution Troubleshooting GuideCharging Indicator Breath Rate Over Capacity Alarm Battery TOO HOT/COLD Alarm Operating Alarm Optional Accessories Invacare XPO2 Limited Warranty 0120