SECTION 6—OPERATING INSTRUCTIONS

Power Supplies

WARNING

Use only Invacare specified power supplies with the XPO2 portable concentrator. Use of other non approved power supplies with the XPO2 can cause damage and/or injury and will void the warranty.

NOTE: For this procedure, refer to FIGURE 6.1.

The XPO2 can be supplied power from a factory installed internal battery, from an AC power adapter connected to the wall outlet, from a DC power adapter connected to a DC outlet like that found in an automobile, or from a supplemental battery pack.

Internal Battery: A factory installed, rechargeable, internal battery is located inside the XPO2. When fully charged, it supplies power for up to 3.5 hours. An audible alarm sounds when the battery power is getting low. Refer to Indicators, Warnings and Alarms on page 33.

AC Power Adapter: An AC power adapter allows the XPO2 to be connected to a 100‐240 volt 50/60 hertz outlet. The power adapter converts AC voltage to a DC voltage that can be used to power the XPO2. Use of the AC power adapter will allow the XPO2 to be operated and simultaneously recharge the

internal battery. This AC power adapter can also be used to recharge the supplemental battery pack. Refer to Charging the Batteries on page 22.

Part No 1148112

19

Invacare XPO2

Image 19
Contents Invacare XPO2 Portable Concentrator Accessories Warning Table of Contents Table of Contents Special Notes Disposal of Equipment and Accessories General Guidelines Operating Information Radio Frequency Interference MaintenancePolarized Plug Instruction Storage Packaging and HandlingUnpacking InspectionTechnical Description XPO2 Features XPO2FEATURESIPX1 Typical Product ParametersBattery is charging while unit is Running Regulatory Listing Filters IntroductionRecommendations for Optimal Performance LocationPower Supplies Power Supplies Checking Charge Level of Internal Battery Checking Charge Level of Internal BatteryCharging the Batteries Charging the Internal BatteryCharging the Supplemental Battery Only Charging the Supplemental Battery Only Operating Instructions Prolonging XPO2 Battery Life Battery Time ManagementDon’ts Powering the XPO2 Powering the XPO2 Connecting/Positioning the Nasal Cannula Connecting/Positioning the Nasal Cannula Adjusting the Pulse Flow SettingAdjusting the Pulse Flow Setting Using the XPO2 Portable ConcentratorIndicators, Warnings and Alarms Operating Instructions Operating Instructions Operating Instructions Cabinet CLEANING, CARE, and MaintenanceAir Intake Filter Air Intake FilterCarrying Bag Carrying BagALARM/WARNING Symptom Probable Cause Solution Troubleshooting GuideCharging Indicator Breath Rate Over Capacity Alarm Battery TOO HOT/COLD Alarm Operating Alarm Optional Accessories Invacare XPO2 Limited Warranty 0120