Invacare XPO100B Using the XPO2 Portable Concentrator, Adjusting the Pulse Flow Setting

Page 32

SECTION 6—OPERATING INSTRUCTIONS

FLOW SETTING

AVERAGE PULSE VOLUME (AT 20

 

 

BREATHS PER MINUTE)

 

 

 

1

 

15 ml

 

 

 

2

 

23 ml

 

 

 

3

 

31 ml

 

 

 

4

 

37 ml

 

 

 

5

 

42 ml

 

 

 

Indicator Lights

On/Off Button

Flow Selection

FIGURE 6.9 Adjusting the Pulse Flow Setting

Using the XPO2 Portable Concentrator

1.Turn the XPO2 On by pressing the On/Off button.

2.Breathe normally through your nose. Breathing through your mouth may result in less effective oxygen therapy. The XPO2 will put out a measured pulse of oxygen each time it detects inhalation. The blue flow setting indicator light will blink each time an inhalation is detected and the XPO2 outputs a pulse of oxygen.

3.Turn the XPO2 Off by pressing and holding the power On/Off button for approximately two seconds.

NOTE: The time required for the XPO2 to reach maximum oxygen concentration after being turned On is approximately 5 minutes.

Invacare XPO2

32

Part No 1148112

Image 32
Contents Invacare XPO2 Portable Concentrator Accessories Warning Table of Contents Table of Contents Special Notes Disposal of Equipment and Accessories General Guidelines Operating Information Maintenance Radio Frequency InterferencePolarized Plug Instruction Packaging and Handling UnpackingInspection StorageTechnical Description XPO2FEATURES XPO2 FeaturesTypical Product Parameters IPX1Battery is charging while unit is Running Regulatory Listing Introduction FiltersLocation Recommendations for Optimal PerformancePower Supplies Power Supplies Checking Charge Level of Internal Battery Checking Charge Level of Internal BatteryCharging the Internal Battery Charging the BatteriesCharging the Supplemental Battery Only Charging the Supplemental Battery Only Operating Instructions Battery Time Management Prolonging XPO2 Battery LifeDon’ts Powering the XPO2 Powering the XPO2 Connecting/Positioning the Nasal Cannula Adjusting the Pulse Flow Setting Connecting/Positioning the Nasal CannulaUsing the XPO2 Portable Concentrator Adjusting the Pulse Flow SettingIndicators, Warnings and Alarms Operating Instructions Operating Instructions Operating Instructions CLEANING, CARE, and Maintenance CabinetAir Intake Filter Air Intake FilterCarrying Bag Carrying BagTroubleshooting Guide ALARM/WARNING Symptom Probable Cause SolutionCharging Indicator Breath Rate Over Capacity Alarm Battery TOO HOT/COLD Alarm Operating Alarm Optional Accessories Invacare XPO2 Limited Warranty 0120