SECTION
Air Intake Filter
CAUTION
DO NOT operate the XPO2 without the air intake filter installed.
NOTE: For this procedure, refer to FIGURE 7.1.
NOTE: Remove the air intake filter and clean it at least once each week.
1.Lift filter cover slightly and pull down to remove tabs from the grooves.
2.Lift out the filter.
3.Use a vacuum cleaner or wash with a mild liquid dish detergent (such as Dawn™) and water. Rinse thoroughly.
4.Thoroughly dry the filter and inspect for fraying, crumbling, tears and holes. Replace filter if any is found.
5.Reinstall the filter cover by placing the tabs in the slots and engaging the magnet.
NOTE: Use only Invacare part number 1148281 as the air intake filter for your XPO2.
Magnet | Cover | Grooves | Cover | Tabs |
|
|
|
|
Magnet
Filter
FIGURE 7.1 Air Intake Filter
Invacare XPO2™ | 38 | Part No 1148112 |