SECTION 6—OPERATING INSTRUCTIONS

Battery Discharged Alarm – Once the XPO2 internal battery

(or supplemental battery, if applicable) capacity drops to minimum, a triple audible beep will sound and the GREEN 25% battery gauge indicator will blink very rapidly. The unit will then shutdown.

No Breath Detect Alarm ‐ When the XPO2 is operating but does not sense breathing, a constant audible alarm sounds and the red alarm indicator light will illuminate continuously. If this occurs, check the connection from the cannula to the XPO2 unit and ensure that you are breathing through your nose. If the alarm continues, change to an alternate source of oxygen and contact your equipment provider.

Breath Rate Over Capacity Alarm – If your breathing rate causes the capacity of the XPO2 to be exceeded, an intermittent audible beep sounds and the red alarm indicator will illuminate intermittently. When this occurs, reduce your activity and/or change to another source of oxygen.

If your breathing rate continues to increase the audible beeps and red indicator blinks will become more rapid.

NOTE: Oxygen will be supplied during this alarm.

System Too Hot/Cold for Start Alarm – If the internal sensors register temperatures outside factory set levels upon start‐up, the unit will alarm with a rapid audible beep, the red alarm indicator light will illuminate continuously, the unit will not operate and the flow setting 1 & 2 blue indicator lights will illuminate. The fan will turn On. Allow the unit to cool or warm to the recommended operating temperature range (see Recommendations for Optimal Performance on page 18). Turn Off the unit and try again. Change to an alternate source of oxygen while waiting. If the alarm continues, contact your equipment provider.

Invacare XPO2

34

Part No 1148112

Image 34
Contents Invacare XPO2 Portable Concentrator Accessories Warning Table of Contents Table of Contents Special Notes Disposal of Equipment and Accessories General Guidelines Operating Information Maintenance Radio Frequency InterferencePolarized Plug Instruction Inspection Packaging and HandlingUnpacking StorageTechnical Description XPO2FEATURES XPO2 FeaturesTypical Product Parameters IPX1Battery is charging while unit is Running Regulatory Listing Introduction FiltersLocation Recommendations for Optimal PerformancePower Supplies Power Supplies Checking Charge Level of Internal Battery Checking Charge Level of Internal BatteryCharging the Internal Battery Charging the BatteriesCharging the Supplemental Battery Only Charging the Supplemental Battery Only Operating Instructions Battery Time Management Prolonging XPO2 Battery LifeDon’ts Powering the XPO2 Powering the XPO2 Connecting/Positioning the Nasal Cannula Adjusting the Pulse Flow Setting Connecting/Positioning the Nasal CannulaUsing the XPO2 Portable Concentrator Adjusting the Pulse Flow SettingIndicators, Warnings and Alarms Operating Instructions Operating Instructions Operating Instructions CLEANING, CARE, and Maintenance CabinetAir Intake Filter Air Intake FilterCarrying Bag Carrying BagTroubleshooting Guide ALARM/WARNING Symptom Probable Cause SolutionCharging Indicator Breath Rate Over Capacity Alarm Battery TOO HOT/COLD Alarm Operating Alarm Optional Accessories Invacare XPO2 Limited Warranty 0120