ResMed ST, AUTO 25 manual Data Card Card Error Remove Card

Page 17

Problem/Possible Cause

Solution

Displays error message: Check tube!! Key if done

The air tubing is loose or blocked.

Check that the air tubing is connected securely to your mask

 

and the air outlet. Press the Start/Stop key to restart the

device. If this does not clear the message, disconnect the power cord and then reconnect it to restart the device.

Displays error message: Exxxx Call Service (where xxxx defines an error)

Component failure.

Record error number and contact your ResMed service center.

Displays error message: HIGH LEAK!!! Adjust Mask

You have experienced excessively high leak levels for more than 20 seconds.

Check that your air tubing is connected properly. Adjust headgear.

The following message is displayed on the LCD after you try to update settings or copy data to

the Data Card: Card Error Remove Card

Data Card is not inserted correctly.

You may have removed the Data Card before settings were copied to the VPAP device.

Ensure that the Data Card is inserted with the arrow facing up as far as it can go.

Reinsert the Data Card and wait for the Settings Success Remove Card or Copy Complete Remove Card message to appear on the LCD.

The following message is displayed on the LCD after you try to update the settings using the Data Card: Settings Invalid Remove Card

The identification details on the

Contact your clinician/service provider immediately.

Data Card do not match the

details on your device.

The following message is displayed on the LCD after you try to update the settings using the Data Card: Settings Error Remove Card

There is a data error on the Data

Contact your clinician/service provider immediately.

Card.

 

The following message is NOT displayed on the LCD after you try to update the settings using the Data Card: Settings Success Remove Card

The settings were not updated.

Contact your clinician/service provider immediately.

14

Image 17
Contents English User GuideVpap ST Contents Vpap System IntroductionContraindications Adverse EffectsResScan Data Card MasksHumidifier Supplemental Oxygen Using the Vpap Auto 25, Vpap ST & Vpap SSetting up your Vpap Attaching a H4i Humidifier How to Use the Control PanelDown RightVAuto mode CPAP, S, ST & T modes Using the MenusSmartStart How to Select the Mask TypeExample Starting TreatmentStopping Treatment Mask-fit star rating Reminders on the Vpap LCDCopying Data onto a Data Card Using the Data CardTraveling with the Vpap Auto 25, Vpap ST & Vpap S Daily Cleaning and MaintenanceWeekly MonthlyServicing Problem/Possible Cause Solution TroubleshootingData Card Card Error Remove Card Technical Specifications Technical SpecificationsSymbols Which Appear On The Device General Warnings and Cautions Separation distance according to frequency of transmitter m General Warnings and CautionsPage Limited Warranty

AUTO 25, ST, S specifications

The ResMed S, ST, and AUTO 25 devices represent a significant advancement in sleep therapy technology, designed to enhance the treatment of patients with obstructive sleep apnea (OSA) and other sleep-related disorders. These devices are part of ResMed's commitment to improving patient outcomes through innovative breathing solutions.

One of the standout features of the ResMed S, ST, and AUTO 25 is their adaptive pressure support capabilities. The devices utilize advanced algorithms to automatically adjust air pressure based on the user’s breathing patterns. This real-time responsiveness ensures that patients receive optimal therapy tailored to their specific needs, allowing for a more comfortable and effective treatment experience.

The automatic titration feature of the AUTO 25 is particularly noteworthy. It continuously monitors the patient’s airflow and adjusts pressure levels throughout the night. This dynamic capability not only enhances compliance but also helps to prevent apnea events more effectively than fixed-pressure devices.

In terms of comfort, the ResMed S, ST, and AUTO 25 feature a ramp function that gradually increases pressure from a lower starting point. This allows patients to fall asleep more comfortably before reaching the prescribed therapeutic pressure. The quiet operation of these devices adds to patient comfort, ensuring a peaceful night’s sleep for both the user and their partner.

Connectivity is another key characteristic of the ResMed S, ST, and AUTO 25 devices. Equipped with wireless technology, these devices allow for seamless data transfer to healthcare providers. This connectivity facilitates remote monitoring, enabling clinicians to track therapy progress and make necessary adjustments without the need for in-person visits.

Additionally, the user-friendly interface simplifies setup and navigation. Intuitive controls and a clear display make it easy for users to monitor their therapy progress and understand their treatment journey. The devices are designed with patient education in mind, providing users with essential information about their sleep patterns and therapy effectiveness.

ResMed also places a strong emphasis on the durability and reliability of its devices. The S, ST, and AUTO 25 are constructed with high-quality materials and are rigorously tested to ensure long-lasting performance. Maintenance is straightforward, with easily accessible components to facilitate cleaning and upkeep.

In summary, the ResMed S, ST, and AUTO 25 devices represent a comprehensive solution for sleep therapy, combining innovative technologies, user-friendly features, and reliable performance to improve patient care. Their intelligent adaptability, advanced monitoring, and enhanced comfort make them a preferred choice for individuals seeking effective treatment for sleep disorders. With ResMed’s ongoing commitment to research and development, patients can expect even greater advancements in sleep therapy solutions in the future.