Omega OSP SERIES manual Log

Page 33

Press the <MEM> key to start the acquisition.

If automatic acquisition mode is set, OSP acquires automatically with the programmed time interval.

If MANual mode is set, the real time temperature measurement is displayed; press the <MEM> key to store the displayed value.

If MAX or MIN mode is set, the maximum (minimum) temperature measurement is displayed; press the <MEM> key to trigger acquisition and the <ENTER> key to reset the value for the next acquisition.

Press the <SEL> key to obtain the page that allows the clearing of the memory stored data :

Log

°C

Lock

e

Press the <ENTER> key to clear the memory content and to obtain the following message that require confirmation :

Log

°C

Lock

e

Press the <ENTER> key to confirm the cancellation request and to obtain the following message :

Log

°C

Lock

e

Press the <SEL> key to obtain the following page indicating the identification code (or name) assigned to the active Tag (zzzzzzz = Furn686 = Trap888). The identification code is selectable only through software.

Log

°C

Lock

e

Press the <SEL> key to review the pages relevant with -HAL- and -LAL- alarm settings. To modify the alarm level in the data acquisition mode see the paragraph 10.2.

33

Image 33
Contents OSP Series Portable IR thermometers OMEGAnetSM On-Line Service Internet e-mail Info@omega.com Introductory Note Table of Contents Data Acquisition Operative Mode General features General DescriptionInnovative design Instrument codes Thermocouple measuring ranges SpecificationsGeneral General FeaturesOptical System Target pinpointingDisplay KeyboardDigital Interface Self calibrationCalculated Measurements Logging ModeCase Logging Data ManagerPhysical Description Functional Description Power supplyBattery charger MicrocontrollerFirmware Unpacking ΛEM = 10.6 ∝M PRE-OPERATIONAL CheckHow to maximize the life span of the battery Power SupplyPower supply with alkaline batteries Power supply from main line ACAnalogue input Laser SightPage Operations How to Operate the instrumentQuick Start Trigger * MEM SEL EnterLaser sighting HoldWorking band Weight Unstable temperature measurementAlarm settings HAL Minimum, maximum, average and difference indication Automatic Emissivity Setting Thermocouple input enabled ConfigurationTemperature Scale selection Technical Unit selectionAcquisition settings Date & Real Time clock settingBuzzer On/Off Ambient Temperature Compensation TAMOSP 500/800 decimal point selection Actual Target Emissivity Ambient Temp Indicated Temp TAMInstrument serial number Firmware revision numberBattery level of charge Data Acquisition Operative Mode How to operate in data acquisition modeAcquisition by dedicated Tags Log Recall stored data Installation Logging Data ManagerProgram Architecture Help FileData EditToolbars Copy into an open file Quick startingMove into a new file Move into an open filePress the or key to select the Tag a identification code Log Page Options & Accessories Printer operationsSighting telescope system Installation of the sighting device Filter installationRed Point pinpointing system Part n DescriptionPage Infrared energy Application NotesApplications EmissivityDigital output wiring practice Digital InterfaceTTL to RS 232 adapter Communication protocol Computer data request from OSP to PCComputer Reading values $00 Pyrometer measureWriting values Computer data setting from PC to OSPExample a Data transfer from OSP to PC Communication programsExample B PC instructions to OSP Example a9Ø Chksum = DATA1 + DATA2 + DATA3 + DATA4 and &H7F Example BFaulty operating conditions MaintenanceStorage ErrAppendix A1.1 Typical Emissivity Values A1 How to determine an object emissivity14 µm A1.2 Metals Typical Emissivity ValuesA1.3 Non-metals Typical Emissivity Values Index Weight Working band WARRANTY/DISCLAIMER Temperature