Omega OSP SERIES Operations, Quick Start, How to Operate the instrument, Trigger * MEM SEL Enter

Page 20
Keep pressed for operation. Press twice, in sequence, to hold the instrument -ON- Value setting or data selection during Configuration or Set-up procedures
Laser sighting on/off or data acquisition mode set-up Data acquisition operation and logging manual step Selects the operative mode
Memory load key

9 OPERATIONS

OSP series portable infrared thermometers are factory calibrated before shipment.

During the start-up the operator should only select and load, if required, the pertinent application parameter as described in the following paragraphs.

Before entering the procedure to operate the instrument it is useful the understanding of the messages that can be present in the display.

The following figure shows the overall layout of the display and an explanation of each symbol or message :

 

2

 

3

 

4

 

7

1

 

 

 

 

 

 

5

15

Log

Rcl

Set

+ -

°F

 

Auto

 

 

 

K

°C

6

 

 

 

 

 

 

 

 

 

 

 

8

13

 

 

 

 

 

 

9

 

Lock

Comp

HAL

LAL

14

 

 

 

 

 

 

 

10

 

 

 

 

 

 

 

12

 

 

e

 

 

 

11

 

 

 

 

 

 

1 - Indicates logging mode operative

2 - Enables recall of memory stored data

3- Indicates the setting mode enabled during configuration and logging mode set up

4- Low battery indication

5- Temperature in degree Fahrenheit

6- Temperature in degree Celsius

7- Temperature in Kelvin

8- Actual measured temperature value

9- Indicates Low Alarm

10- Indicates High Alarm

11- Auxiliary measurement data or setting (AVG, LAL, HAL, MIN, MAX, DIF) or thermocouple input indication

12- Operative emissivity value

13- Ambient temperature compensation

14- Continuous measurement mode

15- Automatic acquisition in logging mode

The contact closure of the keys is acknowledged as a coded signal by the microprocessor that recognizes the operator’s instructions.

<Trigger> <▲> <▼> <*> <MEM> <SEL> <ENTER>

9.1Quick Start

To use the OSP infrared thermometer right away follow this simple steps :

Point the instrument at the object you want to measure and pull the trigger (the actual temperature value, the pre- programmed emissivity value and the auxiliary indication will be displayed); a “LOCK” symbol will appear on display.

Set emissivity value using the <▲> and <▼> keys. Refer to appendix A to define emissivity and for emissivity tables. Press and release the trigger to switch the unit off.

9.2How to Operate the instrument

To switch the instrument “On” keep the trigger pressed; when the trigger is released the instrument will be switched “Off”. When continuous operations are required press twice, in sequence, the trigger to lock the operative mode.

The first display indication will show the type of instrument and the temperature technical unit enabled.

20

Image 20
Contents OSP Series Portable IR thermometers OMEGAnetSM On-Line Service Internet e-mail Info@omega.com Introductory Note Table of Contents Data Acquisition Operative Mode Innovative design General featuresGeneral Description Instrument codes Specifications Thermocouple measuring rangesGeneral Features GeneralOptical System Target pinpointingKeyboard DisplayDigital Interface Self calibrationLogging Mode Calculated MeasurementsCase Logging Data ManagerPhysical Description Power supply Functional DescriptionFirmware Battery chargerMicrocontroller Unpacking PRE-OPERATIONAL Check ΛEM = 10.6 ∝MPower Supply How to maximize the life span of the batteryPower supply with alkaline batteries Power supply from main line ACLaser Sight Analogue inputPage How to Operate the instrument OperationsQuick Start Trigger * MEM SEL EnterHold Laser sightingUnstable temperature measurement Working band WeightAlarm settings HAL Minimum, maximum, average and difference indication Automatic Emissivity Setting Configuration Thermocouple input enabledTemperature Scale selection Technical Unit selectionDate & Real Time clock setting Acquisition settingsAmbient Temperature Compensation TAM Buzzer On/OffOSP 500/800 decimal point selection Actual Target Emissivity Ambient Temp Indicated Temp TAMFirmware revision number Instrument serial numberBattery level of charge Acquisition by dedicated Tags Data Acquisition Operative ModeHow to operate in data acquisition mode Log Recall stored data Program Architecture InstallationLogging Data Manager File HelpData EditToolbars Quick starting Copy into an open fileMove into a new file Move into an open filePress the or key to select the Tag a identification code Log Page Sighting telescope system Options & AccessoriesPrinter operations Filter installation Installation of the sighting deviceRed Point pinpointing system Part n DescriptionPage Application Notes Infrared energyApplications EmissivityTTL to RS 232 adapter Digital output wiring practiceDigital Interface Computer Communication protocolComputer data request from OSP to PC $00 Pyrometer measure Reading valuesComputer data setting from PC to OSP Writing valuesCommunication programs Example a Data transfer from OSP to PCExample B PC instructions to OSP Example aExample B 9Ø Chksum = DATA1 + DATA2 + DATA3 + DATA4 and &H7FMaintenance Faulty operating conditionsStorage ErrAppendix A1 How to determine an object emissivity A1.1 Typical Emissivity ValuesA1.2 Metals Typical Emissivity Values 14 µmA1.3 Non-metals Typical Emissivity Values Index Weight Working band WARRANTY/DISCLAIMER Temperature