Omega OSP SERIES manual Microcontroller, Firmware, Battery charger

Page 14

4.3Microcontroller

The microcontroller handles all the logic functions of the instrument, performs the linearization for non linear transducers, compensates for the reference junction temperature, drives the digital display and acknowledges all the operator’s instructions.

The core of the circuit is a single-chip microcomputer that utilizes HCMOS technology to provide the low power characteristics and high noise immunity of CMOS plus the high speed operation of HMOS.

The microcomputer provides highly sophisticated, on- chip peripheral functions including: 256 bytes of static RAM, an 8 channel analog to digital (A/D) converter (used to read the Rj value, the setting of the input comparator, the battery package voltage and the value of the auxiliary input), a serial communication interface (SCI) subsystem, and a serial peripheral interface (SPI) subsystem.

The microprocessor works with an 8-bit communication bus to EPROM and EEPROM memories and is interfaced with a decoder, a latch of address and an inverter-driver.

4.4Firmware

The operating system firmware handles all the logic instructions to the internal peripheral circuits and performs the computation of the linearization equations.

The application system firmware is resident on the non-volatile memory (EEPROM) of the microprocessor chip. It is used to store the installation parameters (autocalibration data, program data, etc.)

4.5Display

The custom display, placed on an auxiliary board, uses high contrast LCD technologies (STN liquid). OSP thermometers are standard equipped with a backlight device for easy readings in poor light conditions.

4.6Battery charger

The auxiliary module, supplied as a standard accessory, allows operations from 115 Vac 50/60 Hz. OSP, if needed, can be operated directly from a line source through the charger.

The plastic case of the battery charger incorporates the line voltage plug and a cable with a connector for interconnections to the instrument.

The charger circuit is designed with an insulating transformer and a voltage stabilizer circuit. The step-down transformer reduces the power line 115 Vac to a value of 10 Vac. The above voltage is full wave rectified , filtered and stabilized. The output voltage of 5.6 Vdc is the ideal value to recharge the internal Ni-MH batteries

4.7Digital interface

The digital interface circuit is essentially based on the serial communication interface subsystem (SCI) on the chip of the microprocessor at 0 / +5V level. Adapters to convert TTL to RS 232 voltage levels are available from OMEGA.

14

Image 14
Contents OSP Series Portable IR thermometers OMEGAnetSM On-Line Service Internet e-mail Info@omega.com Introductory Note Table of Contents Data Acquisition Operative Mode Innovative design General featuresGeneral Description Instrument codes Specifications Thermocouple measuring rangesOptical System General FeaturesGeneral Target pinpointingDigital Interface KeyboardDisplay Self calibrationCase Logging ModeCalculated Measurements Logging Data ManagerPhysical Description Power supply Functional DescriptionFirmware Battery chargerMicrocontroller Unpacking PRE-OPERATIONAL Check ΛEM = 10.6 ∝MPower supply with alkaline batteries Power SupplyHow to maximize the life span of the battery Power supply from main line ACLaser Sight Analogue inputPage Quick Start How to Operate the instrumentOperations Trigger * MEM SEL EnterHold Laser sightingUnstable temperature measurement Working band WeightAlarm settings HAL Minimum, maximum, average and difference indication Automatic Emissivity Setting Temperature Scale selection ConfigurationThermocouple input enabled Technical Unit selectionDate & Real Time clock setting Acquisition settingsOSP 500/800 decimal point selection Ambient Temperature Compensation TAMBuzzer On/Off Actual Target Emissivity Ambient Temp Indicated Temp TAMFirmware revision number Instrument serial numberBattery level of charge Acquisition by dedicated Tags Data Acquisition Operative ModeHow to operate in data acquisition mode Log Recall stored data Program Architecture InstallationLogging Data Manager Data FileHelp EditToolbars Move into a new file Quick startingCopy into an open file Move into an open filePress the or key to select the Tag a identification code Log Page Sighting telescope system Options & AccessoriesPrinter operations Red Point pinpointing system Filter installationInstallation of the sighting device Part n DescriptionPage Applications Application NotesInfrared energy EmissivityTTL to RS 232 adapter Digital output wiring practiceDigital Interface Computer Communication protocolComputer data request from OSP to PC $00 Pyrometer measure Reading valuesComputer data setting from PC to OSP Writing valuesExample B PC instructions to OSP Communication programsExample a Data transfer from OSP to PC Example aExample B 9Ø Chksum = DATA1 + DATA2 + DATA3 + DATA4 and &H7FStorage MaintenanceFaulty operating conditions ErrAppendix A1 How to determine an object emissivity A1.1 Typical Emissivity ValuesA1.2 Metals Typical Emissivity Values 14 µmA1.3 Non-metals Typical Emissivity Values Index Weight Working band WARRANTY/DISCLAIMER Temperature