Invacare IRC5LXO2AWQ, IRC5LXAW Recommended Guidelines for Optimum Performance, Plug in Power Cord

Page 20

SECTION 5—OPERATING INSTRUCTIONS

Recommended Guidelines for Optimum Performance

Temperature:

(10°C - 35°C) 50°F - 95°F

 

 

Electrical:

No extension cords.

 

 

Placement:

No closer than 7.5 cm (3 in) from the wall,

 

furniture, draperies, or similar surfaces.

 

 

Tubing and Cannula:

IRC5LXAW, IRC5LXO2AW,

 

IRC5LXO2AWQ

 

2.1 m (7 ft) cannula with a maximum 15.2 m

 

(50 ft) of Crush-Proof Tubing (DO NOT

 

pinch).

 

 

 

IRC9LXO2AWQ

 

Recommended use up to 15.2 m (50 ft) high

 

flow tubing with high flow cannula at all flow

 

rates.

 

 

Environment:

Smoke and soot-free. No confined spaces

 

(Example: No closets).

 

 

Relative Humidity:

20 to 60%

 

 

Time of Operation:

Up to 24 hours per day.

 

 

Flow Rate

IRC5LXAW, IRC5LXO2AW,

 

IRC5LXO2AWQ

 

0.5 L/min. to 5 L/min. Flow rates less than 1

 

L/min., are not recommended.

 

 

 

IRC9LXO2AWQ

 

1 L/min. to 9L/min. Flow rates less than 1

 

L/min. are not recommended.

 

 

Minimum Operating Time:

30 Minutes

NOTE: Ensure that your concentrator is at least 7.5 cm (3 in) away from walls, draperies or furniture to assure sufficient air flow. Avoid deep pile carpets and heaters, radiators or hot air registers.

Plug in Power Cord

1. Plug in power cord to an electrical outlet.

Platinum™

20

Part No 1134866

Image 20
Contents Platinum Accessories Table of Contents Troubleshooting Guide Limited Warranty Special Notes Recycling Information General Guidelines General Guidelines Maintenance Radio Frequency InterferenceOxygen Concentration ImmediatelyFeatures Front ViewShipping and Handling UnpackingInspection StorageUnpacking Typical Product Parameters ListingIRC5LXO2AW Flow Alarm IRC5LXAW/IRC5LXO2AW Status Indicator Lights Select a Location Operating InstructionsIntroduction Plug in Power Cord Recommended Guidelines for Optimum PerformanceConnect Humidifier If So Prescribed Filling the HumidifierAttaching the Humidifier Bottle Adapter Power Switch Attaching the Oxygen TubingPower Switch FlowrateSensO2 Oxygen Purity Indicators LX O2 Sensor Model Initial Startup of the ConcentratorIndicators LX Model Explanation of LX models Indicator LightsMain Power Loss Alarm System Elapsed Time Meter Hour MeterMaintenance Routine MaintenanceCabinet Preventive Maintenance Record Troubleshooting Guide SolutionTroubleshooting Guide Symptom Limited Warranty Limited Warranty Limited Warranty 0434