SECTION
Recommended Guidelines for Optimum Performance
Temperature: | (10°C - 35°C) 50°F - 95°F |
|
|
Electrical: | No extension cords. |
|
|
Placement: | No closer than 7.5 cm (3 in) from the wall, |
| furniture, draperies, or similar surfaces. |
|
|
Tubing and Cannula: | IRC5LXAW, IRC5LXO2AW, |
| IRC5LXO2AWQ |
| 2.1 m (7 ft) cannula with a maximum 15.2 m |
| (50 ft) of |
| pinch). |
|
|
| IRC9LXO2AWQ |
| Recommended use up to 15.2 m (50 ft) high |
| flow tubing with high flow cannula at all flow |
| rates. |
|
|
Environment: | Smoke and |
| (Example: No closets). |
|
|
Relative Humidity: | 20 to 60% |
|
|
Time of Operation: | Up to 24 hours per day. |
|
|
Flow Rate | IRC5LXAW, IRC5LXO2AW, |
| IRC5LXO2AWQ |
| 0.5 L/min. to 5 L/min. Flow rates less than 1 |
| L/min., are not recommended. |
|
|
| IRC9LXO2AWQ |
| 1 L/min. to 9L/min. Flow rates less than 1 |
| L/min. are not recommended. |
|
|
Minimum Operating Time: | 30 Minutes |
NOTE: Ensure that your concentrator is at least 7.5 cm (3 in) away from walls, draperies or furniture to assure sufficient air flow. Avoid deep pile carpets and heaters, radiators or hot air registers.
Plug in Power Cord
1. Plug in power cord to an electrical outlet.
Platinum™ | 20 | Part No 1134866 |