Platinum™
30
Part No 1134866
FIGURE 6.2 Preventive Maintenance Record
PREVENTIVE MAINTENANCE RECORD
INVACARE PLATINUM OXYGEN CONCENTRATOR
ON EACH INSPECTION
Record Date of Service
Record Elapsed Hours On Hour Meter
Clean Cabinet Filter(s)
Check Prescribed L/min. Flow Rate
EVERY 4,380 HOURS FOR IRC5LXA W
Check Oxygen Concentration
(UNITS WITHOUT SensO 2 )
EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST
Check Oxygen Concentration
(UNITS WITH SensO 2 )
DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
Clean/Replace Cabinet Filter(s)
Check Outlet HEPA Filter*
Check Compressor Inlet Filter*
Check Power Loss Alarm
AS REQUIRED
Clean Heat Exchanger
Replace Exhaust Muffler
Rebuild Top End of Compressor
*NOTE: Refer to Preventive Maintenance section of Service Manual. NOTE:
2,160 hours are equivalent to usage 24 hours per day, for 90 days.
4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months. 26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years.
MODEL NO. I RC
SERIAL NO.
SECTION