SECTION 5—OPERATING INSTRUCTIONS

3.Remove the filter access door located on the side of the concentrator. Insert a flathead screwdriver in the plate groove on the top edge of the filter access door and gently pry the filter access door open. Refer to Detail “A” in FIGURE 5.3.

4.The humidifier adapter is next to the inlet filter. Pull up and remove the humidifier bottle adapter. Refer to Detail “B” in FIGURE 5.3.

5.Attach it to the humidifier by turning the wing nut on the humidifier bottle counterclockwise until it is securely fastened. Refer to Detail “C” in FIGURE 5.3.

DETAIL “A”

Plate Groove (Insert Flat

Head Screwdriver Here)

Filter Access Door

DETAIL “B”

Humidifier Bottle

Adapter

DETAIL “C”

Filter

Access

DoorHumidifier Bottle

Adapter

FIGURE 5.3 Attaching the Humidifier Bottle Adapter

Platinum™

22

Part No 1134866

Image 22
Contents Platinum Accessories Table of Contents Troubleshooting Guide Limited Warranty Special Notes Recycling Information General Guidelines General Guidelines Maintenance Radio Frequency InterferenceOxygen Concentration ImmediatelyFeatures Front ViewInspection Shipping and HandlingUnpacking StorageUnpacking Typical Product Parameters ListingIRC5LXO2AW Flow Alarm IRC5LXAW/IRC5LXO2AW Status Indicator Lights Introduction Operating InstructionsSelect a Location Plug in Power Cord Recommended Guidelines for Optimum PerformanceConnect Humidifier If So Prescribed Filling the HumidifierAttaching the Humidifier Bottle Adapter Power Switch Attaching the Oxygen TubingPower Switch FlowrateSensO2 Oxygen Purity Indicators LX O2 Sensor Model Initial Startup of the ConcentratorIndicators LX Model Explanation of LX models Indicator LightsMain Power Loss Alarm System Elapsed Time Meter Hour MeterMaintenance Routine MaintenanceCabinet Preventive Maintenance Record Troubleshooting Guide SolutionTroubleshooting Guide Symptom Limited Warranty Limited Warranty Limited Warranty 0434