Invacare IRC5LXO2AWQ, IRC9LXO2AWQ, IRC5LXAW manual Flow Alarm

Page 16

SECTION 4—TYPICAL PRODUCT PARAMETERS

Change in maximum

0.7 L/min.

recommended flow when

 

back pressure of 7k Pa is

 

applied:

 

 

 

Filters:

Cabinet (2), Outlet HEPA and Compressor

 

Inlet.

 

 

Safety System:

Current overload or line surge shutdown.

 

High temperature compressor shutdown.

 

High Pressure Alarm w/compressor

 

shutdown.

 

Low Pressure Alarm w/compressor

 

shutdown.

 

Battery Free Power Loss Alarm.

 

SensO2 Oxygen System (SensO2 Model) Low

 

Flow Alarm.

 

 

Width:

46.7 cm ± 1 cm (18 3/8 in ± 3/8 in)

 

 

Height:

67.0 cm ± 1 cm (6 3/8 in. ± 3/8 in)

 

 

Depth:

36.5 cm ± 1 cm (14 3/8 in. ± 3/8 in)

 

 

Weight:

IRC5LXAW, IRC5LXO2AW,

 

IRC5LXO2AWQ

 

23.6 kg ± 1 kg (52 lbs ± 2 lbs)

 

 

 

IRC9LXO2AWQ

 

24.0 kg ± 1 kg (53 lbs ± 2 lbs)

 

 

Shipping Weight:

IRC5LXAW, IRC5LXO2AW,

 

IRC5LXO2AWQ

 

26.8 kg ± 1 kg (59 lbs ± 2 lbs)

 

 

 

IRC9LXO2AWQ

 

27.2 kg ± 1 kg (60 lbs ± 2 lbs)

 

 

Operating Ambient

10°C - 35°C (50°F - 95°F) at 20-60% relative

Temperature:

humidity.

 

 

Exhaust:

Less than Ambient +19°C (+ 35°F)

 

 

Oxygen Output:

Less than Ambient +2°C (+ 4°F).

 

 

Cabinet:

Impact Resistant flame-retardant plastic

 

cabinet that conforms to UL 94-V0.

 

 

Platinum™

16

Part No 1134866

Image 16
Contents Platinum Accessories Table of Contents Troubleshooting Guide Limited Warranty Special Notes Recycling Information General Guidelines General Guidelines Maintenance Radio Frequency InterferenceOxygen Concentration ImmediatelyFeatures Front ViewShipping and Handling UnpackingInspection StorageUnpacking Typical Product Parameters ListingIRC5LXO2AW Flow Alarm IRC5LXAW/IRC5LXO2AW Status Indicator Lights Introduction Operating InstructionsSelect a Location Plug in Power Cord Recommended Guidelines for Optimum PerformanceConnect Humidifier If So Prescribed Filling the HumidifierAttaching the Humidifier Bottle Adapter Power Switch Attaching the Oxygen TubingPower Switch FlowrateSensO2 Oxygen Purity Indicators LX O2 Sensor Model Initial Startup of the ConcentratorIndicators LX Model Explanation of LX models Indicator LightsMain Power Loss Alarm System Elapsed Time Meter Hour MeterMaintenance Routine MaintenanceCabinet Preventive Maintenance Record Troubleshooting Guide SolutionTroubleshooting Guide Symptom Limited Warranty Limited Warranty Limited Warranty 0434