SECTION
NOTE: The use of some accessories such as the pediatric flowstand and the HomeFill compressor will deactivate the Low Flow Alarm.
Flow Knob
Ball
FIGURE 5.6 Flowrate
SensO2 Oxygen Purity Indicators (LX O2 Sensor Model)
This feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate.
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 min.) while waiting for the O2 purity to reach maximum.
When the unit is turned on, the GREEN light will come on (SYSTEM OK/O2 greater than 85%). After five minutes, the oxygen sensor will be operating normally and will control the indicator lights depending on oxygen concentration values.
Explanation of Oxygen Purity Indicator Lights
NOTE: For this procedure, refer to FIGURE 5.7.
Part No 1134866 | 25 | Platinum™ |