SECTION
SECTION 4—TECHNICAL
DESCRIPTION
The Invacare Perfecto2 concentrator is used by patients with
respiratory disorders who require supplemental oxygen. The device is not intended to sustain or support life.
The oxygen concentration level of the output gas ranges from 87% to 95.6%. The oxygen is delivered to the patient through the use of a nasal cannula.
The Invacare Perfecto2 concentrator uses a molecular sieve and pressure swing adsorption methodology to produce the oxygen gas output. Ambient air enters the device, is filtered and then compressed. This compressed air is then directed toward one of two nitrogen adsorbing sieve beds. Concentrated oxygen exits the opposite end of the active sieve bed and is directed into an oxygen reservoir where it is delivered to the patient.
The Invacare Perfecto2 concentrator is capable of operation by the patient in a home environment or in an institutional environment. Device operates at a nominal 230 VAC/50 Hertz supply.
Service information will be available upon request to qualified technical personnel only.
Perfecto2™Series | 12 | Part No 1150713 |