SECTION 8—TROUBLESHOOTING GUIDE
SYMPTOM: | PROBABLE CAUSE: |
| SOLUTION: | |
|
|
|
| |
Alarm: | Main Power Loss: |
|
| |
|
|
|
|
|
Short beeps, long | 1. | Power cord not | 1. | Insert plug into outlet. |
pause | plugged in. |
|
| |
|
|
|
|
|
Concentrator not | 2. | No power at outlet. | 2. | Inspect house circuit |
operating, |
|
| breakers or fuses. If problem | |
power switch On. |
|
| recurs, use a different outlet. | |
|
|
|
|
|
Beep......Beep | 3. | Tripped circuit | 3. | Push/reset circuit breaker. |
| breaker. | If problem recurs, call Service | ||
|
|
| Provider. | |
|
|
| ||
Alarm: | System Failure: | 1a. Remove and clean cabinet | ||
Continuous | 1. | Unit overheating due | filters. | |
| to blocked air intake. |
|
| |
Concentrator not |
|
| 1b. Move oxygen | |
operating, Power | 2. | Insufficient power at | concentrator at least three | |
Switch On. | outlet. | inches away from walls, | ||
|
|
| draperies or furniture. | |
Beeeeeeeep.... | 3. | Internal repairs |
|
|
| required. | 2. | DO NOT use extension | |
|
|
| cords. Move to another | |
|
|
| electrical outlet or circuit. | |
|
|
| 3. | Call Service Provider. |
|
|
|
|
|