Invacare IRC5PO2AW, IRC5PAW manual Troubleshooting Guide Symptom

Page 28

SECTION 8—TROUBLESHOOTING GUIDE

SYMPTOM:

PROBABLE CAUSE:

SOLUTION:

 

 

 

YELLOW or RED

1. Low oxygen purity.*

1. Clean or Replace filters.

light Illuminated.

 

 

 

 

 

 

2. Kinked or blocked

2. Inspect for kinks or

 

tubing, cannula or

blockages. Correct, clean or

Alarm:

humidifier.*

replace item. Once

 

corrected, turn power Off for

Continuous

 

 

60 seconds and then turn

On RED light

 

 

power back On.

only.

 

 

 

 

3. Flowmeter set at 0.5

3. Check flowmeter is set to

 

L/min *.

1.0 L/min or more. Refer to

 

 

Typical Product Parameters

* Only applies to

 

on page 13.

IRC5PO2AW

 

 

4. Unit overheating due

4a. Remove and clean cabinet

models.

to blocked air intake.

filters.

 

 

 

4b. Move concentrator at

 

 

least three inches from walls,

 

 

draperies, and furniture.

 

 

 

 

5. Insufficient power at

5. DO NOT use extension

 

outlet.

cords. Move to another

 

 

electrical outlet or circuit.

 

 

 

 

6. Internal repairs

6. Call Service Provider.

 

required.

 

 

 

 

GREEN light with

1. Internal repairs

1. Call Service Provider.

YELLOW Light

required.

 

Flashing.

 

 

 

 

 

Alarm:

LOW FLOW ALARM

1a. Inspect for kinks or

Rapid

1a. Kinked or blocked

blockages. Correct, clean or

 

tubing, cannula or

replace item. Once

Beep....Beep...

humidifier.

corrected, turn power Off for

Beep...Beep

 

60 seconds and then turn

 

1b. Flowmeter set at 0.5

power back ON.

 

L/min.

 

 

 

1b. Flowrates less than 1

 

 

L/min are not recommend.

 

 

 

Perfecto2™Series

28

Part No 1150713

Image 28
Contents Perfecto2Series Accessories Accessories WarningTable of Contents Signal Word Meaning Special NotesSpecial Notes Label Location Perfecto2General Guidelines Operating Information Radio Frequency Interference Features Handling UnpackingInspection StorageTechnical Description Typical Product Parameters IPX0IRCPF16AW IRC5PO2AW/IRC5PAW O2 Purity Indicator Lights Instructions OperatingIntroduction Select a LocationSet Up Filling the HumidifierHumidifer Bottle Adapter Power Switch Humidifier CompartmentFlowrate FlowrateSensO2 Oxygen Purity Indicator Model IRC5PO2AW Initial Startup of the ConcentratorIRC5PAW Explanation of Indicator Lights ModelElapsed Time Meter Hour Meter Maintenance Routine MaintenanceCleaning the Cabinet Model No. IRC Serial No Troubleshooting Guide SolutionTroubleshooting Guide Symptom Optional Accessories Recycling Information Limited Warranty 0120