Invacare IRC5PO2AW, IRC5PAW manual IRCPF16AW

Page 14

SECTION 5—TYPICAL PRODUCT PARAMETERS

Flow Range:

0.5 to 5 L/min (maximum). For flowrates less

 

than 1 L/min, we recommend the use of the

 

Invacare Pediatric Flowmeter Accessory

 

(IRCPF16AW).

 

 

Low Flow Alarm:

0 L/min to 0.5 L/min

 

Rapid Audible Alarm Beeping (No

 

Accessories Connected).

 

 

 

 

Average Power

300 W

 

 

Consumption:

280 W @

>

3 L/min

 

 

 

Pressure Relief Mechanism

241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)

Operational at:

 

 

 

 

 

 

 

Change in maximum

0.7 L/min

 

 

recommended flow when

 

 

 

back pressure of 7kPa is

 

 

 

applied:

 

 

 

 

 

Filters:

Cabinet, Outlet HEPA and Compressor Inlet

 

 

Safety System:

Current overload or line surge shutdown. High

 

temperature compressor shutdown. High

 

Pressure Alarm w/compressor shutdown. Low

 

Pressure Alarm w/compressor shutdown.

 

Battery Free Power Loss Alarm. SensO2 Oxygen

 

System (SensO2 Model) Low Flow Alarm.

 

 

Width:

38.1 cm ± 1 cm (15 in ± 3/8 in)

 

 

Height:

58.4 cm ± 1 cm (23 in ± 3/8 in)

 

 

Depth:

30.5 cm ± 1 cm (12 in ± 3/8 in)

 

 

Weight:

20.5 kg ± 1 kg (45 lbs ± 2 lbs)

 

 

Shipping Weight:

22.7 kg ± 1 kg (50 lbs ± 2 lbs)

 

 

Operating Ambient

10°C - 35°C (50°F - 95°F) at 20-60% relative

Temperature:

humidity

 

 

 

 

Exhaust Cooling Air

Less than Ambient +19°C (+ 35°F)

Temperature:

 

 

 

 

 

Oxygen Output

Less than Ambient +3°C(+ 6°F)

Temperature:

 

 

 

 

 

Cabinet:

Impact Resistant flame-retardant plastic cabinet

 

that conforms to UL 94-V0.

 

 

 

 

 

 

 

 

Perfecto2™Series

14

Part No 1150713

Image 14
Contents Perfecto2Series Accessories Accessories WarningTable of Contents Signal Word Meaning Special NotesSpecial Notes Label Location Perfecto2General Guidelines Operating Information Radio Frequency Interference Features Inspection HandlingUnpacking StorageTechnical Description Typical Product Parameters IPX0IRCPF16AW IRC5PO2AW/IRC5PAW O2 Purity Indicator Lights Introduction InstructionsOperating Select a LocationSet Up Filling the HumidifierHumidifer Bottle Adapter Power Switch Humidifier CompartmentFlowrate FlowrateSensO2 Oxygen Purity Indicator Model IRC5PO2AW Initial Startup of the ConcentratorElapsed Time Meter Hour Meter Explanation of Indicator Lights ModelIRC5PAW Maintenance Routine MaintenanceCleaning the Cabinet Model No. IRC Serial No Troubleshooting Guide SolutionTroubleshooting Guide Symptom Optional Accessories Recycling Information Limited Warranty 0120