SECTION 6—OPERATING INSTRUCTIONS

Your oxygen concentrator will perform best when operated under the conditions outlined in the Typical Product Parameters on page 13. Usage in environments other than those described may result in the need for increased equipment maintenance. The air intake of the unit should be located in a well ventilated area to avoid airborne pollutants and/or fumes.

Set Up

1.Plug in power cord to an electrical outlet.

2.Connect Humidifier (if so prescribed).

WARNING

DO NOT overfill humidifier.

DO NOT reverse the oxygen input and output connections. Water from the humidifier bottle will travel through the cannula back to the patient.

NOTE: For this procedure, refer to FIGURE 6.1, FIGURE 6.2 on page 19, FIGURE 6.3 on page 20.

1.Remove cap from bottle.

2.Fill humidifier with distilled water to the level indicated by the manufacturer. Replace the humidifier cap and securely tighten.

Humidifier

HumidifierBottle w/o Cap

Bottle with Cap

FIGURE 6.1 Filling the Humidifier

Perfecto2™Series

18

Part No 1150713

Image 18
Contents Perfecto2Series Accessories Accessories WarningTable of Contents Signal Word Meaning Special NotesSpecial Notes Label Location Perfecto2General Guidelines Operating Information Radio Frequency Interference Features Inspection HandlingUnpacking StorageTechnical Description Typical Product Parameters IPX0IRCPF16AW IRC5PO2AW/IRC5PAW O2 Purity Indicator Lights Introduction InstructionsOperating Select a LocationSet Up Filling the HumidifierHumidifer Bottle Adapter Power Switch Humidifier CompartmentFlowrate FlowrateSensO2 Oxygen Purity Indicator Model IRC5PO2AW Initial Startup of the ConcentratorExplanation of Indicator Lights Model IRC5PAWElapsed Time Meter Hour Meter Maintenance Routine MaintenanceCleaning the Cabinet Model No. IRC Serial No Troubleshooting Guide SolutionTroubleshooting Guide Symptom Optional Accessories Recycling Information Limited Warranty 0120