Invacare HF2E540M9 manual Inspection and Compressor Interface, Cylinder Prefill Inspection

Page 10

SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

NOTE: Refer to the HomeFill ® Compressor Owner’s Manual for compressor operating instructions.

Cylinder Prefill Inspection

WARNING

All cylinders MUST be inspected before attempting to fill; otherwise, injury or damage may occur.

NEVER smoke in an area where oxygen is being administered.

NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere.

NOTE: For this procedure, refer to on page 12.

The Hydrostatic Testing Date

Aluminum cylinders MUST undergo testing every five years.

WARNING

DO NOT fill cylinders that have not been tested in the past five years. Otherwise, serious personal injury may result. Contact your dealer for replacement.

HomeFill® Cylinder

10

Part No 1163156

Image 10
Contents HomeFill Cylinder Oxygen 540 Valve HomeFill Cylinder Table of Contents Receiving Inspection Special NotesIntended USE General Guidelines General Guidelines General Guidelines Features Product DiagramTypical Product Parameters Inspection and Compressor Interface Cylinder Prefill InspectionExternal Examination Checking Cylinder Pressure Filling the Cylinder Fillport Covers Resetting Connector Fillport Connecting the Cylinder to the Compressor Disconnecting the Cylinder from the Compressor Connector Fillport Maintenance Troubleshooting Cylinder Fill TimesHomeFill Cylinder Limited Warranty Limited Warranty USA