SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

External Examination

1.Examine the outside of the cylinder for the following conditions and replace the cylinder if they exist:

Dents or dings

Arc burns

Oil or grease

Any other signs of damage that might cause a cylinder to be unacceptable or unsafe for use.

2.Examine the cylinder for evidence of fire or thermal damage. Evidence includes charring or blistering of the paint, or other protective coating or heat sensitive indicator. If fire or thermal damage is found, replace the cylinder.

3.Inspect the cylinder fillport for the following and replace if found:

Debris, oil or grease

Noticeable signs of damage

Signs of corrosion inside the valve

Signs of excessive heat or fire damage

Part No 1163156

11

HomeFill® Cylinder

Image 11
Contents HomeFill Cylinder Oxygen 540 Valve HomeFill Cylinder Table of Contents Intended USE Special NotesReceiving Inspection General Guidelines General Guidelines General Guidelines Product Diagram FeaturesTypical Product Parameters Cylinder Prefill Inspection Inspection and Compressor InterfaceExternal Examination Checking Cylinder Pressure Filling the Cylinder Fillport Covers Resetting Connector Fillport Connecting the Cylinder to the Compressor Disconnecting the Cylinder from the Compressor Connector Fillport Maintenance Cylinder Fill Times TroubleshootingHomeFill Cylinder Limited Warranty Limited Warranty USA