SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

WARNING

The fillport cover on the connector fillport and the post valve fillport MUST be replaced after filling and whenever not in use. If either of the fillport covers are missing contact your healthcare provider before using.

17.Place the fillport covers onto the connector fillport (located on the compressor) and post valve fillport (located on the cylinder) (FIGURE 4.6).

18.Install a 540 post valve compatible conserver or regulator according to the manufacturer’s instructions.

19.Open the cylinder valve by turning counterclockwise and examine the conserver or regulator gauge to ensure cylinder is full (needle pointing into GREEN area). If the cylinder is not full, perform the following:

Repeat Filling the Cylinder on page 13.

If cylinder is not full after second attempt to fill, contact the provider.

Cylinder

Valve

Fillport Cover

Connector

Fillport Cover

 

FIGURE 4.6 Connector Fillport

HomeFill® Cylinder

18

Part No 1163156

Image 18
Contents HomeFill Cylinder Oxygen 540 Valve HomeFill Cylinder Table of Contents Special Notes Receiving InspectionIntended USE General Guidelines General Guidelines General Guidelines Features Product DiagramTypical Product Parameters Inspection and Compressor Interface Cylinder Prefill InspectionExternal Examination Checking Cylinder Pressure Filling the Cylinder Fillport Covers Resetting Connector Fillport Connecting the Cylinder to the Compressor Disconnecting the Cylinder from the Compressor Connector Fillport Maintenance Troubleshooting Cylinder Fill TimesHomeFill Cylinder Limited Warranty Limited Warranty USA