Invacare HF2E540M9 manual Filling the Cylinder

Page 13

SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

Filling the Cylinder

NOTE: For this procedure, refer to FIGURE 4.1 thru FIGURE 4.6.

1.Check the cylinder pressure before attempting to fill cylinder. Refer to Checking Cylinder Pressure on page 12.

2.Close the cylinder valve by rotating clockwise (FIGURE 4.1).

3.Remove the conserver or regulator according to the device manufacturer instructions (FIGURE 4.1).

CONSERVER/REGULATOR

INSTALLED

Conserver/

Cylinder Regulator

Valve

CONSERVER/REGULATOR

REMOVED

Gauge

Cylinder

FIGURE 4.1 Component Identification

4.Remove the post valve cover (located on the cylinder) and the connector fillport covers (located on the compressor) (FIGURE 4.2).

Part No 1163156

13

HomeFill® Cylinder

Image 13
Contents HomeFill Cylinder Oxygen 540 Valve HomeFill Cylinder Table of Contents Receiving Inspection Special NotesIntended USE General Guidelines General Guidelines General Guidelines Product Diagram FeaturesTypical Product Parameters Cylinder Prefill Inspection Inspection and Compressor InterfaceExternal Examination Checking Cylinder Pressure Filling the Cylinder Fillport Covers Resetting Connector Fillport Connecting the Cylinder to the Compressor Disconnecting the Cylinder from the Compressor Connector Fillport Maintenance Cylinder Fill Times TroubleshootingHomeFill Cylinder Limited Warranty Limited Warranty USA