Page 15

SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

NOTE: If the outer ring (sleeve) is in the UP position (GREEN dots not visible), the connector fillport will not be able to accept the cylinder fillport. Pushing DOWN momentarily will reset the connector fillport (GREEN dots visible) to accept the cylinder fillport.

PUSH

Outer Ring (Sleeve)

DOWN

Connector Fillport

GREEN Dots

 

BEFORE

COUPLING

CYLINDER,

PUSH DOWN

ON SLEEVE

UNTIL

GREEN DOTS

ARE VISIBLE.

FIGURE 4.3 Resetting Connector Fillport

6.Grasp the cylinder assembly as shown in Detail “A” in FIGURE 4.4.

7.Position the cylinder in the compressor cradle as shown in Detail “C” in FIGURE 4.4.

8.Align the cylinder fillport with the connector fillport as shown in Detail “C” in FIGURE 4.4.

9.Pull UP on the outer ring (sleeve) of the connector fillport (Detail “B”) while pushing DOWN on the cylinder assembly to couple the cylinder fillport into the connector fillport as shown in Detail “B” in FIGURE 4.4.

NOTE: The cylinder is properly connected when an audible “click” is heard.

10. Press the compressor power switch to the On ( ) position.

Part No 1163156

15

HomeFill® Cylinder

Image 15
Contents HomeFill Cylinder Oxygen 540 Valve HomeFill Cylinder Table of Contents Special Notes Receiving InspectionIntended USE General Guidelines General Guidelines General Guidelines Product Diagram FeaturesTypical Product Parameters Cylinder Prefill Inspection Inspection and Compressor InterfaceExternal Examination Checking Cylinder Pressure Filling the Cylinder Fillport Covers Resetting Connector Fillport Connecting the Cylinder to the Compressor Disconnecting the Cylinder from the Compressor Connector Fillport Maintenance Cylinder Fill Times TroubleshootingHomeFill Cylinder Limited Warranty Limited Warranty USA