SECTION 6—CYLINDER FILL TIMES
NOTE: All filling times are approximate and may vary on environmental conditions.
*
Cylinder Filling | Concentrator Flow Rate to |
Times | Patient: |
| (Platinum 5 up to 2.5 L/min and |
| Platinum 10 up to 5 L/min) |
|
|
NORMAL | 2 hr 20 min |
|
|
MAXIMUM | 2 hr 45 min |
|
|
Troubleshooting
If post valve cylinder fails to function, try the following solutions. If problem cannot be corrected, consult your dealer/home care equipment provider.
PROBLEM | POSSIBLE CAUSE |
| SOLUTION | |
|
|
|
|
|
No flow. | 1. | Regulator or | 1. | Set to prescribed |
| conserver is in “Off” | setting. | ||
| position. |
|
| |
| 2. | Cylinder empty. | 2. | Replace/fill cylinder. |
| 3. | Cylinder Valve | 3. | Open toggle valve. |
| closed. |
|
| |
| 4. | Conserver not | 4a. Check position of | |
| sensing breath. | cannula in nose. Do not | ||
|
|
| breathe through mouth. | |
|
|
| 4b. Check to see if the | |
|
|
| cannula is kinked. | |
|
|
|
|
|
Replacement Parts - Accessories
HF2POST9BAG Carrying Bag.