MaxTech OM-25ME operating instructions External Probe

Page 12

13

 

 

 

and Keys- The and arrow keys are used in

 

 

 

 

conjunction with the CALIBRATE key to calibrate the unit or in

 

conjunction with the LOW SET and HIGH SET keys to adjust the alarm

 

settings. When used after pressing the CALIBRATE key, pressing the

 

or key will raise or lower the displayed oxygen value in .1%

 

increments. When either of these keys are held down for more than 1

 

second, the display will scroll at a rate of .4% per second. When used

 

after pressing either the LOW SET or HIGH SET keys, pressing the or

 

key will raise or lower the displayed alarm setting in 1% increments. In

 

the alarm set mode, holding these keys down for more than 1 second

 

will cause the display to scroll at a rate of 4% per second.

14

 

 

 

LOW SET Key- When the keypad is unlocked, pressing the

 

 

 

 

 

 

 

 

 

 

 

 

LOW SET key will cause the "LOW" icon to flash. The low

 

 

 

 

 

alarm setting can then be changed using the and arrow keys. When

 

the low alarm value is set and the LOW SET or LOCK/UNLOCK key is

 

pressed again, the unit exits the alarm set mode and returns to normal

 

operation. The unit will also revert to normal operation (storing the most

 

recent alarm value) if 10 seconds elapse and no keys are pressed. The

 

OM-25ME alarm setting is adjustable down to 18%. The OM-25ME-L low

 

alarm setting is adjustable to 15%.

15

 

 

 

HIGH SET Key- When the keypad is unlocked, pressing the

 

 

 

 

HIGH SET key will cause the "HIGH" icon to flash. The high

 

alarm setting can then be changed using the and arrow keys. When

 

the high alarm value is set and the HIGH SET or LOCK/UNLOCK key is

 

pressed again, the unit exits the alarm set mode and returns to normal

 

operation. The unit will also revert to normal operation (storing the most

 

recent alarm value) if 10 seconds elapse and no keys are pressed.

16

 

 

SILENT Key- In an alarm condition, pressing the SILENT key will

 

 

 

deactivate the alarm beeper, but the alarm LED and display will

 

continue to flash. In 120 seconds, the alarm beeper will sound again.

 

This process will be repeated until the alarm condition is cancelled. For

 

an alarm condition to be cancelled, the concentration must be .1% higher

 

than the alarm setting (in the presence of a low alarm condition) or .1%

 

lower than the alarm setting (in the presence of a high alarm condition).

D External Probe

17

Sensor with Diverter- The sensor (with diverter) is designed to fit

industry standard, 15 mm I.D. "T" adapters.

18

Extendible Cable- The extendible cable allows the sensor to be

positioned up to 10 feet from the side of the unit.

5

Image 12
Contents Model OM-25ME Page Page Preface Iii Page SET-UP Procedure Base Unit Specifications Sensor SpecificationsBase Unit Description Components Description please refer to LCD ReadoutLCD Readout Alarm LEDs Keypad External ProbeAlarm LEDs External Probe MAX-250E Oxygen Sensor Battery Installation Calibrating the MAXO2 Monitor Before You BeginTo Calibrate the MAXO2 Monitor Automatic Calibration to Room Air Effects of Temperature Pressure Effect Humidity EffectAlarm Setting Procedure Low Alarm Setting Basic Operation Sensor Removal and Replacement Problem Solving Base Unit Specifications Exposure to Anesthetic Gases Calibration Techniques in Pressurized SystemsCalibration Errors Page Spare Parts and Accessories Warranty Page Email sales@maxtecinc.com