MaxTech OM-25ME operating instructions Base Unit Specifications, Sensor Specifications

Page 23

7.

SPECIFICATIONS

 

7.1

Base Unit Specifications

 

Measurement Range:

0.0-100%

 

Resolution:

0.1%

 

Accuracy and Linearity: ±1% of full scale at constant temperature, R.H.

 

 

and pressure when calibrated at full scale.

Total Accuracy:

±3% Actual Oxygen Level over full operating

 

 

temperature range.

 

Response Time:

90% of final value in approximately 15 seconds

 

 

at 23 C

 

Warm-up Time:

none required

 

Operating Temperature: 15 C - 40 C (59 F - 104 F)

Storage Temperature:

-15 C - 50 C (5 F - 122 F)

Humidity:

0-95% (non-condensing)

Power Requirements:

2, AA Alkaline batteries (2 X 1.5 Volts)

Battery Life:

approximately 3000 hours in typical use.

Low Battery Indication:

"LOW BAT" icon displayed on LCD

Sensor Type:

Maxtec MAX-250E galvanic fuel cell

Expected Sensor Life:

>900,000% O2 Hours

 

 

 

over 2 years in typical applications

Alarm System:

high/low alarms, flashing red LEDs, nominal

 

 

2300Hz audible alarm beeper

Low Alarm Range:

18%-99% (>1% lower than high alarm)

 

 

15%-99% (>1% lower than high alarm)OM-25ME-L

High Alarm Range:

19%-99% (>1% higher than low alarm)

 

 

16%-99% (>1% higher than low alarm)OM-25ME-L

Alarm Accuracy:

exact to displayed alarm value

Dimensions:

3.5"(W) x 5.5"(H) x 1.5"(D) [89mm x 140mm x 38mm]

Weight:

approximately .92 lbs. (417g)

Cable Length:

10 ft. (3m) fully extended

Diverter Fitting:

fits industry standard, 15 mm "T" adapter

7.2

Sensor Specifications

 

Type:

galvanic fuel sensor (0-100%)

Life:

 

2 years in typical applications

Interferent

Volume % Dry

Interference in O2%

Nitrous Oxide

75%

<2%

Carbon Dioxide

10%

<2%

Halothane

5%

<2%

Enflurane

5%

<2%

Isoflurane

5%

<2%

Helium

70%

<2%

Sevoflurane

6%

<2%

Desflurane

15%

<2%

16

Image 23
Contents Model OM-25ME Page Page Preface Iii Page Base Unit Specifications Sensor Specifications SET-UP ProcedureBase Unit Description LCD Readout Components Description please refer toKeypad External Probe LCD Readout Alarm LEDsAlarm LEDs External Probe MAX-250E Oxygen Sensor Calibrating the MAXO2 Monitor Before You Begin Battery InstallationTo Calibrate the MAXO2 Monitor Automatic Calibration to Room Air Effects of Temperature Humidity Effect Pressure EffectAlarm Setting Procedure Low Alarm Setting Basic Operation Sensor Removal and Replacement Problem Solving Base Unit Specifications Calibration Techniques in Pressurized Systems Exposure to Anesthetic GasesCalibration Errors Page Spare Parts and Accessories Warranty Page Email sales@maxtecinc.com