MaxTech OM-25ME operating instructions Calibration Errors

Page 25

8.3 Calibration Errors

The MAXO2 Monitors have a self test feature built into the software to detect faulty calibrations. During calibration, if the signal from the oxygen sensor is outside the limits stored within the instrument’s memory, a flashing “CAL Er” is displayed. The error code is displayed to indicate that either the sensor should be replaced or that there is a fault in the calibration process. A few simple hints can prevent calibration errors.

If you try to adjust the monitor display before the reading has stabilized, the “CAL Er” may appear. For example, if the monitor had just been calibrated on a known high concentration of oxygen source gas and then exposed to ambient air, you should wait until the reading has stabilized. If you try to adjust the display to read 20.9% before the sample line has cleared of high concentration O2, the sensor may actually be exposed to residual high % oxygen. The signal from the sensor would still be high and considered “out of spec” for air, thus resulting in a “CAL Er”. The proper procedure is to wait for the reading to stabilize before adjusting the display. This may take 30 seconds or more.

Externally mounted sensors, as found on the OM-25MEor OM-25ME-Lcome equipped with diverter tips. The tips help direct the gas in a “T” fitting up to the sensor for analysis. The diverter tips should be only used with a flowing gas. When calibrating in a non-flowing environment, remove the diverter tip.

8.4 Atmospheres of High Humidity

The MAXO2 Monitor can be used in applications where the relative humidity of the sample gas ranges from 0 to 95%, non-condensing. However, it should be noted that water vapor exerts its own pressure in the same manner as oxygen does in a sample gas stream.

For example, if the monitor is calibrated in dry gas and then the gas is humidified, the monitor will correctly display a reading which is slightly lower than previously displayed. This is due to the dilution of oxygen in the sample gas by water vapor.

This fact is important to note in systems where there exist both “wet” and “dry” gas streams such as in a ventilator circuit. If the monitor is measuring oxygen on the “dry side” of the ventilator, it will correctly indicate an oxygen concentration slightly greater than actually found in the “wet side” (delivered to the patient). The water vapor has diluted the gas stream.

18

Image 25
Contents Model OM-25ME Page Page Preface Iii Page Base Unit Specifications Sensor Specifications SET-UP ProcedureBase Unit Description LCD Readout Components Description please refer toKeypad External Probe LCD Readout Alarm LEDsAlarm LEDs External Probe MAX-250E Oxygen Sensor Calibrating the MAXO2 Monitor Before You Begin Battery InstallationTo Calibrate the MAXO2 Monitor Automatic Calibration to Room Air Effects of Temperature Humidity Effect Pressure EffectAlarm Setting Procedure Low Alarm Setting Basic Operation Sensor Removal and Replacement Problem Solving Base Unit Specifications Calibration Techniques in Pressurized Systems Exposure to Anesthetic GasesCalibration Errors Page Spare Parts and Accessories Warranty Page Email sales@maxtecinc.com