MaxTech OM-25ME operating instructions Sensor Removal and Replacement

Page 21

4. SENSOR REMOVAL AND REPLACEMENT

The OM-25ME is shipped with a new MAX-250E oxygen sensor installed. Although the sensor has a very long expected life, eventually the sensor will require replacement. Removing or installing a sensor, when necessary, is a very simple procedure. To remove and install a new sensor:

1)Grasp the sensor in one hand and, with the other hand, unscrew the cable connector counter-clockwise at the sensor.

2)Pull out the cable connector plug from the expired sensor.

3)Unscrew the flow diverter from the sensor and discard the expired sensor.

Note: The sensor contains lead and lead acetate, be sure to dispose of expired sensors in accordance with hospital, local, state and federal regulations.

4)Remove the new sensor from the packaging and remove the protective film from the sensor face.

5)Insert the cable connector plug into the receptacle of the new sensor and tighten the cable connector.

6)Screw the flow diverter onto the new sensor.

7)Wait approximately 20 minutes for the sensor to reach equilibrium.

8)Calibrate the new sensor.

Note: If the monitor is on when the sensor is detached and replaced, the monitor will automatically force a re-calibration. The display will read “CAL”.

14

Image 21
Contents Model OM-25ME Page Page Preface Iii Page Base Unit Specifications Sensor Specifications SET-UP ProcedureBase Unit Description LCD Readout Components Description please refer toKeypad External Probe LCD Readout Alarm LEDsAlarm LEDs External Probe MAX-250E Oxygen Sensor Calibrating the MAXO2 Monitor Before You Begin Battery InstallationTo Calibrate the MAXO2 Monitor Automatic Calibration to Room Air Effects of Temperature Humidity Effect Pressure EffectAlarm Setting Procedure Low Alarm Setting Basic Operation Sensor Removal and Replacement Problem Solving Base Unit Specifications Calibration Techniques in Pressurized Systems Exposure to Anesthetic GasesCalibration Errors Page Spare Parts and Accessories Warranty Page Email sales@maxtecinc.com