MaxTech OM-25ME operating instructions Components Description please refer to, LCD Readout

Page 9

1.2Components Description (please refer to page 3)

A 1

LCD Readout

3 1/2-Digit Display- The 3 1/2-digit liquid crystal display (LCD) provides direct readout of oxygen concentrations in the range of 0-100%. It also displays "CAL" when the calibration mode is entered.

2

"%" Sign- In the calibration mode, the "%" sign flashes every

 

second.

3

Low Battery Indicator- The low battery indicator appears on

 

the LCD display when the power supply voltage drops below

 

acceptable limits. When the "LOW BAT" icon is visible, batteries

 

should be replaced immediately.

4

LOCKED Status Indicator- The presence of the "LOCKED"

 

icon on the LCD readout is an indication that the unit is in its

 

normal "LOCKED" state. The unit must be unlocked, using the

 

LOCK key, in order to be calibrated or to change the high and low

 

alarm settings.

5

Low Alarm Indicator- The low alarm setting is displayed at all

 

times just below the "LOW" icon on the LCD readout. The

 

indicated value represents the oxygen percentage at which the

 

low alarm will be activated. In the low alarm set mode, the

 

"LOW" icon will flash at a 1 second rate. In a low alarm

 

condition, the low alarm value will flash in conjunction with the

 

low alarm light emitting diode (LED).

6

High Alarm Indicator- The high alarm setting is displayed at all

 

times just below the "HIGH" icon on the LCD readout. The

 

indicated value represents the oxygen percentage at which the

 

high alarm will be activated. In the high alarm set mode, the

 

"HIGH" icon will flash at a 1 second rate. In a high alarm

 

condition, the high alarm value will flash in conjunction with the

 

high alarm LED.

7

Threaded Mounting Bushing- Threads to Table Top Swivel

 

Stand. Also may be used with many third party camera mounting

accessories.

2

Image 9
Contents Model OM-25ME Page Page Preface Iii Page Base Unit Specifications Sensor Specifications SET-UP ProcedureBase Unit Description LCD Readout Components Description please refer toKeypad External Probe LCD Readout Alarm LEDsAlarm LEDs External Probe MAX-250E Oxygen Sensor Calibrating the MAXO2 Monitor Before You Begin Battery InstallationTo Calibrate the MAXO2 Monitor Automatic Calibration to Room Air Effects of Temperature Humidity Effect Pressure EffectAlarm Setting Procedure Low Alarm Setting Basic Operation Sensor Removal and Replacement Problem Solving Base Unit Specifications Calibration Techniques in Pressurized Systems Exposure to Anesthetic GasesCalibration Errors Page Spare Parts and Accessories Warranty Page Email sales@maxtecinc.com