Toshiba LF410/LF602, LF410/LF600 specifications Overall Specifications

Page 1

Field Intelligent Device Series

LF410 /LF600

LF410 /LF602

Electromagnetic Flowmeter

15 to 200 mm (1/2" to 8")

 

 

 

 

Introduction

The electromagnetic flowmeter uses Faraday’s Law of electromagnetic induction to measure the process flow. The device consists of two units: a detector, through which the fluid to be measured flows and in which low-level signals proportional to flow rates are obtained; and a converter, which supplies excitation current to the detector, and amplifies the signals from the detector and then processes and converts the signals into the 4–20mAdc current signal or communication signal. Thanks to the unique patented magnetic field distribution technology, the meter is highly immune for upstream flow disturbances. Combined with multi-functional converter LF600 (combined type) or LF602 (separate type) equipped with its original noise-suppression circuit and arithmetic operation capability, LF410 has high tolerance to noise, giving stable output even for slurry fluid measurement. IR (Infrared) switches enable parameter setting of the converter without removing the cover. Flow direction can be set in either way, and its 128 x 128 dot matrix LCD display allows the LCD rotated to 90, 180 and 270 degrees with a software.

LF410/LF600

LF410

LF602

LF414/LF600

LF414

 

Figure2. LF410 series Flowmeters

*3

Certification number

Z01207

Specifications

Overall Specifications

The AF900 hand-held terminal (HART*1 communicator) can be used to communicate with the flowmeter from a remote place. PROFIBUS-PA*2interface is available as option.

*1: HART protocol (Highway Addressable Remote Transducer) is a communication protocol for industrial sensors recommended by the HCF (HART Communication Foundation).

*2: PROFIBUS is the communication protocol for the factory automation and process automation that PROFIBUS Organization recommends. Instead of analog control with a conventional analog signal (4-20mA), it is one kind of the fieldbus which digitized all signals. Flowmeters support PROFIBUS-PA.

Measurement range in terms of flow velocity:

0 – 0.3 m/s to 0 –10 m/s (0 –1.0 ft/s to 0 – 32.8 ft/s). 0 – 0.1 m/s to 0 – 0.3 m/s (0 – 0.3 ft/s to 0 – 1.0 ft/s) range is available optionally.

Accuracy: See the following graph. Pulse output:

Vs > 0.5 m/s (1.64 ft/s): +/-0.5 % of rate.

Vs < 0.5 m/s (1.64 ft/s): +/-0.3% of rate +/- 1 mm/s (0.039 inch/s).

Current output: plus +/-8 uA (0.05 % of span)

Note: Span = Range in the magmeters.

Converter

Converter

 

I/O

 

 

 

Signal cable

Power

 

supply

 

Terminal box

 

 

Power

 

 

supply

 

I/O

Detector

 

Detector

Error

Absolute figure

(% of rate)

 

5.0

 

 

 

4.0

 

 

 

3.0

 

 

 

2.0

 

 

 

1.0

 

 

 

0.0 0

1

2

10

Flow Velosity (m/s)

Accuracy

Combined type

Separate type

Note: The accuracy above is measured under standard

LF410/LF600

LF410/LF602

operating conditions using the weighing method

LF414/LF600

LF414/LF602

at Toshiba's flow calibration facility.

Figure1. Configuration

*3: CE mark is under application.

 

 

 

 

No. EJL-114B

Image 1
Contents Measurement range in terms of flow velocity Overall SpecificationsModel LF600 and LF602 converters Model LF410 and LF414 DetectorsDigital output DO2 opt Digital output DO1 stdCommunications output : HARTstd Vibration resistance Cable connection ports Cable glandsDimensions InstallationSeparate type detectors LF410 and LF414 Meter size 15mm 1/2 Separate type detectors LF410 and LF414 Meter size 25mm Combined type LF410/LF600 and LF414/LF600 flowmeters External ConnectionsSeparate type LF410/LF602 and LF414/LF602 flowmeters Wiring Precautions Profibus Wiring PrecautionsGrounding of the Combined type and Separate type Piping PrecautionsTo select the meter size Meter SizeCalibration Range PED matrix in each flange connectionAbout establishment environment Ordering InformationSpecification Code Wafer type detector LF410 Series Specification Code for converters ISO9001 and ISO14001 are certified