Toshiba LF410/LF600, LF410/LF602 Meter Size, To select the meter size, Calibration Range

Page 12

LF410/LF600 LF410/LF602

Meter Size

To select the meter size:

See Figure 13 and find meter sizes within the velocity of 0.1 to 10m/s for a specified full-scale (measuring range high limit) flow. Select one that has its full-scale velocity between 1 and 3m/s.

Note: Make sure the full-scale flow rate used for the final planning stage stays within 10m/s in terms of flow velocity.

Flow velocity (m/s)

Flow rate (m3/h)

For SI unit

 

 

Unit: m3/h

Meter size

 

Flow rate

 

(mm)

0.3 m/s

1 m/s

10 m/s

15

0.1908

0.6361

6.361

25

0.5301

1.767

17.67

40

1.357

4.523

45.23

50

2.120

7.067

70.67

80

5.428

18.09

180.9

100

8.482

28.27

282.7

150

19.08

63.61

636.1

200

33.93

113.1

1131

For English unit

 

Unit: gal/min

Meter size

 

Flow rate

 

(inch)

0.98ft/s

3.28ft/s

32.8ft/s

1/2

0.8401

2.801

28.01

1

2.334

7.780

77.80

1 1/2

5.975

19.91

199.1

2

9.334

31.12

311.2

3

23.90

79.65

796.5

4

37.35

124.5

1,245

6

84.01

280.1

2,801

8

149.4

498.0

4,980

Figure 13. Flow Rate and Flow Velocity

Calibration Range

If the calibration range is not specified, the standard range as shown below will be used. If the range is specified, we will use the specified range for calibration.

Table 1. Standard Flow Range

Meter size

Standard flow range

mm (inch)

Flow rate

Flow velocity

Flow rate

(m³/h)

(m/s)

(gal/min)

15 (1/2)

2

3.144

25

25 (1)

6

3.395

75

40 (1 1/2)

15

3.316

175

50 (2)

25

3.537

300

80 (3)

60

3.316

650

100 (4)

100

3.537

1,000

150 (6)

200

3.144

2,500

200 (8)

300

2.653

4,500

Note: The unit of "gal/min" is not exchanged (converted) by "m3/h".

PED matrix in each flange connection.

The following sizes fall under the category for PED in each flange connection when the meter ships to EU. All of them had complied with it from a notified body.

Table 2 PED matrix in each flange connection

Flange standard

Meter size

DIN PN 16 and BS 16

150 to 400mm

(6 to 16 inch)

 

DIN PN 10 and BS 10

250 to 400mm

(10 to 16 inch)

 

ANSI 150 and JIS10K

6 to 16 inch

(150 to 400mm)

 

12

Image 12
Contents Overall Specifications Measurement range in terms of flow velocityModel LF410 and LF414 Detectors Model LF600 and LF602 convertersDigital output DO1 std Digital output DO2 optCommunications output : HARTstd Cable connection ports Cable glands Vibration resistanceInstallation DimensionsSeparate type detectors LF410 and LF414 Meter size 15mm 1/2 Separate type detectors LF410 and LF414 Meter size 25mm External Connections Combined type LF410/LF600 and LF414/LF600 flowmetersSeparate type LF410/LF602 and LF414/LF602 flowmeters Wiring Precautions Wiring Precautions ProfibusPiping Precautions Grounding of the Combined type and Separate typeMeter Size To select the meter sizeCalibration Range PED matrix in each flange connectionOrdering Information About establishment environmentSpecification Code Wafer type detector LF410 Series Specification Code for converters ISO9001 and ISO14001 are certified