Toshiba LF410/LF600 Model LF410 and LF414 Detectors, Model LF600 and LF602 converters

Page 2

LF410/LF600 LF410/LF602

Fluid conductivity: 5µS/cm minimum

Fluid temperature:

–10 to +180°C : Ceramic type (14 to 356 °F)

Note : 120°C (248°F) above is separate type –10 to +120°C : tefolon PFA

(14 to 248°F)

Ambient temperature:

Without cFMus Approval:

-20 to + 60 °C (-4 to 140 °F)

With cFMus Approval:

-10 to +45°C (14 to 113 °F)

Structure: NEMA 4X (IP 67) Watertight

Power consumption:

17W(27VA) or less

19W(29VA) or less (with PROFIBUS)

Conformance to European Community Directives:

EMC directive 89/336/EEC

The low voltage 93/68/EEC

PED 97/23/EC (Note 1)

Note : See table 2 for detail.

Approved hazardous location certifications:

Model: LF414/LF600 and LF414/LF602

cFMus explosion proof:

FM Class I, Division 2, Groups A,B,C, and D. FM Class II, Division 2, Groups E, F and G. FM Class III.

Detector and converter combination:

LF410/LF600: Combined type for standard specification.

LF410/LF602: Separate type for standard specification.

LF414/LF600: Combined type with Ex approval of Class I, Division 2 (cFMus).

LF414/LF602: Separate type with Ex approval of Class I, Division 2 (cFMus).

Model LF410 and LF414 Detectors

Mounting style: Wafer type

Fluid pressure: -0.1 to 2.0MPa

(-15 to 300psi, or -1.0 to 20 bar)

Note: The test pressure before shipping from the factory is equal to twice the nominal pressure rating of the customer specified flange connection during 15 minutes.

Connection flange standards: ANSI 150, ANSI 300, BS10 and 16, DIN PN10 and PN16, JIS10K, JIS16K and JIS20K

Principal materials:

Case — 25 to 100mm (1” to 4”): stainless steel 15, 150, and 200mm (1/2”, 6”, and 8”): carbon steel

Linings — 15 to 100mm (1/2” to 4”): Ceramic tube (std.) & teflon PFA (opt.)

150 and 200mm (6” to 8”): teflon PFA Electrodes — 316L stainless steel (std.) Grounding rings — 316 stainless steel (std.)

Note: See Table 3 for optional materials and other related information.

Measuring tube material — 304 stainless steel (in case of Teflon PFA lining)

Coating — 25 to 100mm (1” to 4”): no coating (stainless steel body).

15, 150, and 200mm (1/2”, 6”, and 8”): phthalic acid resin coating with pearl-gray colored.

Heat shock resistance : for a ceramic tube detector Heating: ΔT ≤ 150 °C/0.5sec (302°F/0.5sec) Cooling: ΔT ≤ 100 °C/0.5sec (212°F/0.5sec)

Note: The above means that the ceramic tube detector

withstands the shock of sudden heating (temperature difference 150°C or less per 0.5seconds) and sudden cooling (temperature difference 100°C or less per 0.5seconds).

Dimensions and weights: See Figures 3 to 8.

Cable connection port: for separate type detectors. Cable glands

LF410: without cFMus approval Provided as standard R(PT) 1/2 male screws.

LF414: with cFMus approval

Not provided

3/4-14NPT male screws are required.

Applicable diameter — 11 to 13mm

(0.433 to 0.512 inch)

Model LF600 and LF602 converters

Input signals

Analog signal — the voltage signal from detector, proportional to process flow rate (for LF602 separate type converter).

Digital input DI (opt.)

Signal type: 20 to 30Vdc voltage signal

Input resistance: 2.7kΩ

Number of inputs: one point

DI function — One of the following functions can be assigned to the optional DI signal.

Range switching — Selects either the higher or lower range in the unidirectional or bidirectional 2-range setting.

Totalizer control — Starts and stops the built-in

2

Image 2
Contents Overall Specifications Measurement range in terms of flow velocityModel LF410 and LF414 Detectors Model LF600 and LF602 convertersCommunications output : HARTstd Digital output DO1 stdDigital output DO2 opt Cable connection ports Cable glands Vibration resistanceInstallation DimensionsSeparate type detectors LF410 and LF414 Meter size 15mm 1/2 Separate type detectors LF410 and LF414 Meter size 25mm External Connections Combined type LF410/LF600 and LF414/LF600 flowmetersSeparate type LF410/LF602 and LF414/LF602 flowmeters Wiring Precautions Wiring Precautions ProfibusPiping Precautions Grounding of the Combined type and Separate typeCalibration Range Meter SizeTo select the meter size PED matrix in each flange connectionOrdering Information About establishment environmentSpecification Code Wafer type detector LF410 Series Specification Code for converters ISO9001 and ISO14001 are certified