IBM M/02-212 III.3 Mounting of temperature controlled measuring devices, Ftk-Cc

Page 12

Please read information concerning installation of the computer system in this Operation Manual.

III.3 Mounting of temperature controlled measuring devices

In this chapter you get information concerning mounting and attachment of the following accessories components:

FTK-CC temperature control device for use of cylinder measuring system in the temperature range -10°C...+90°C (liquid’s temperature control)

KE cooling device in conjunction with FTK-CC temperature control device or ME-CP/PP measuring device to rise the temperature range up to -20°C...+180°C

ME-CP/PP measuring device for use of cone/plate and plate/plate measuring system in the temperature range -10°C...+90°C (liquid’s temperature control)

III.3.1 Mounting of temperature control device FTK-CC with and without cooling device KE

As temperature control device is available optionally:

-FTK-CC

for use of cylinder measuring system in the temperature range -10°C...+90°C (liquid’s temperature control).

The temperature control device “FTK-CC” should be used only in the temperature range -10°C ...+90°C.

The cooling device “KE” is necessary additionally for temperature range of - 20°C...+180°C!

thread joint for mounting flange

hose connection of flow output

Pt 100

thermostating chamber

hose connection of flow input

measuring cup screw fitting

Fig.: 4.: Temperature control device FTK-CC

Mounting of the temperature control device “FTK-CC”

-Turn off the R/S Rheometer with the mains switch “POWER” at the instrument back side.

-When using cooling device KE mount cooling device first (see chapter 3.3.3)

-Set the “FTK-CC” from below on the mounting flange of R/S Rheometer and tighten the thread.

-Fix the hoses of liquid circulation thermostat at the “ FTK-CC ” as described below.

-Connect the FTK-CC´s built-in Pt100 cable with the connector “PT100” at the instrument back side of the R/S Rheometer.

Brookfield Engineering Labs., Inc.

Page 12

M/02-212

Image 12
Contents Brookfield R/S Rheometer Brookfield Engineering Labs., Inc 02-212 Contents Appendix General Description Use of the RheometerMeasuring Principle II. System Configuration II.1 R/S RheometerRheometer main instrument minimum configuration II.2 Measuring Devices II.3 Computer SystemIII. Instrument Installation Hints for height adjustment of the standIII.1 Stand Mounting III.2.1 Temperature Sensor Pt100 III.2 Electrical ConnectionsIII.2.2 AC-Adaptor III.2.3 Printer ConnectionIII.2.4 Computer Connection Use data link cable of Brookfield delivery onlyIII.3 Mounting of temperature controlled measuring devices FTK-CCThermostat connection to temperature control device FTK-CC Thermostat connection to temperature control device FTK-CC III.3.2 Mounting of the measuring device ME-CP/PPBrookfield Engineering Labs., Inc 02-212 Brookfield Engineering Labs., Inc 02-212 III.3.3 Mounting of the cooling device KE MountingIV. Environment, handling, cleaning and maintenance IV.3 CleaningIV.1 Operating environment, Storage IV.2 HandlingIV.4 Maintenance Measuring Systems Measurement directly in the substanceMeasurement with substance in measuring system Depth of immersionMeasurement with temperature control device FTK-CC Turn left Turn right Lower Lift Brookfield Engineering Labs., Inc 02-212 VI. Operation and Menue System Rheometer Ver .02 #xxxxxx 07.05.98 BrookfieldVoltage check VCC15.25 VI.1 Keyboard BrookfieldKey-function dependent to executed operation VI.2 Menu System of R/S Rheometer Zero-point calibrationDefine/edit program Print programs from memoryCONFIGURATION-,emie Set time/date Set RS232 Param Language Menu handlingVI.3 Selecting from Lists Main-menue Run Single Run Program RemoteVI.4 Input of numerical values and alphanumeric texts Select Measuring system CC48 CC45Input Values Val.1/s0100.00 Nr. of MP010 Times Input Values Val.1/s0200.00 Nr. of MP010 Times0100Input Values Val.1/s0200.00 Nr. of MP Times0100 Input Values Val.1/s0290.00 Nr. of MP Times0100VI.5 Menue entries Main menue VI.5.1 MAIN-Menue → Run SingleOutput of MPs to No output device Memory Menu ST Start Prog. Running Wait for 1.MPProgram end Single mode Total time100s Total #MP10 VI.5.2 MAIN-Menue → Run ProgramNo valid program Enter program first Select program Prog xyz Test Prog ÖlOutput of MPs to Printer Memory Menue Prog. running Wait for 1.MP #1VI.5.5 Main menue → Configuration VI.5.3 Main menue → RemoteVI.5.4 Main menue → Utilities Program end Test Total time200s Total #MP40VI.6 Menue entries in the Utilities-Menue VI.6.1 Utilities → Zero CalibrationVI.6.2 Utilities → Edit Programs D1/s=ft Select Prog to edit Prog xyz Test NEW Select Meas. system CC45 CC25 CC14Nr. of steps Select Input-Mode -D1/s Nrpm TauPaRange-Error Range-Error Start-Value min 0.90 max Input Step Nr time sEdit Program name VI.6.3 Utilities → Print Programs VI.6.4 Utilities → Measuring systemsTest Press OK to save Printer not ready OK cont. ST stopVI.6.5 Utilities → Print memory Enter Meas. Sys. #4 Name CP25 tauprom 01.1418 kgammaPress OK to save Name CC25 tauprom 01.1418 kgamma Sel. Output-device Printer RS232VI.6.7 Utilities → Measure temperature VI.6.6 Utilities → Clear memoryVI.7 Menu entries of the Configuration menu VI.7.1 Configuration → Set output modeService, Service2 for service personnel only Select output device No output device Printer RS232VI.7.2 Configuration → MeasCount mode VI.7.3 Configuration → Reset meascountVI.7.4 Configuration → Set Time/Date VI.7.5 Configuration → Set RS232 Parameter New Time/Date 1715 Press OK to saveRS232 4800,8,n,1 Change RS232-Par.? Okyes Stno VI.7.6 Configuration → Language VI.7.7 Configuration → Service, ServiceVI.8 Serial Data transfer via Interface RS232 Baudrate -4800 9600Brookfield Engineering Labs., Inc 02-212 Brookfield Engineering Labs., Inc 02-212 VII.1 Measuring in manual mode VII. MeasurementsMeasurement control Measurement with FTK-CCMeasurement in remote control mode VII.2 Measuring in Remote-modeVIII. Technical data Weight Temperature range Measured or evaluated valuesIX. Guarantee A1Data sheets of standard measuring systems Given ranges are standard values not maximum valuesDG DIN Measuring system C25-1 C25-2 C50-1 C50-2 C75-1 C75-2 Table Plate/Plate measuring systems DINA2 Error messages Range ErrorPrinter Error Range error End value min 0.90 maxZero Calibration Error Error #3 Zero cal. error Please retry cal ST stopBreak torque max! Single mode Total MP Total time Break Speed max! Single mode Total MP Total time Break Speed MAXA3 Pin-layout of the serial data cable Rheometer sidePC Side A4 Requirements to the mains connecting cables General informationUSA and Canada Other countriesA5 Language cross reference EnglishGermanEnglishGermanGerman menue system of RC20 Index

M/02-212 specifications

The IBM M/02-212 is a cutting-edge platform that showcases the company’s dedication to excellence in technology and innovation. Designed for both enterprise performance and flexibility, this model adheres to IBM's high standards for processing power and reliability, making it a standout option for businesses looking to enhance their operations.

One of the main features of the M/02-212 is its powerful multi-core processor, which provides exceptional speed and efficiency. This architecture enables the system to handle multiple tasks simultaneously without compromising performance, making it ideal for data-intensive applications and complex processing tasks. It supports a range of workloads, from traditional enterprise applications to modern cloud-based solutions.

In terms of memory, the M/02-212 offers extensive capacity options, utilizing high-speed RAM technology that ensures swift data access and reduced latency. The combination of powerful processing and ample memory resources allows organizations to execute demanding applications and large-scale databases seamlessly.

The system is designed with advanced security features that are integral for protecting sensitive information. Encryption capabilities, secure boot options, and comprehensive identity management functionalities are built-in, providing a robust defense against potential cyber threats. This is particularly valuable for industries like finance and healthcare, where data integrity is paramount.

Moreover, the IBM M/02-212 leverages state-of-the-art virtualization technologies, enabling companies to optimize resource utilization and increase operational efficiency. Through virtualization, multiple virtual machines can run on a single physical server, allowing businesses to reduce hardware costs and improve overall system management.

The M/02-212 also emphasizes scalability, which means organizations can easily expand their capabilities as their needs evolve. This flexibility allows businesses to adapt to changing market conditions without requiring a complete overhaul of their existing infrastructure.

Additionally, it comes equipped with a user-friendly management interface, providing administrators with the tools necessary to monitor system performance, allocate resources, and troubleshoot issues efficiently.

In summary, the IBM M/02-212 combines powerful processing capabilities, robust security measures, advanced virtualization, and exceptional scalability. These characteristics make it an excellent choice for enterprises looking to enhance their IT infrastructure while ensuring efficiency and data security.