IBM M/02-212 operating instructions III.3.3 Mounting of the cooling device KE

Page 17

III.3.3 Mounting of the cooling device KE

The optionally supplied cooling device “KE” must be used in case of the R/S Rheometer operation with the liquid temperature control device “FTK-CC” or measuring device “ME-CP/PP“ in the temperature range below -10°C and above +90°C. When using cooling device KE temperature range of temperature control device “FTK-CC“ and measuring device “ME-CP/PP“ is expanded to -20°C... +180°C.

The cooling liquid flows through the cooling channel of the cooling device KE and prevents the heat transfer from the liquid’s temperated chamber of temperature control device “FTK-CC” or measuring device “ME-CP/PP“ to the R/S Rheometer.

Mounting

-Turn off the R/S Rheometer with the mains switch “POWER” at the instrument back side.

-Lift the “KE” from below to the R/S Rheometer and tighten the thread.

-Fix the hoses of the cooling loop at the “KE” (see below “Cooling water connection to cooling device KE”).

Fig. 6: cooling device KE

Cooling water connection to cooling device KE

The hoses of cooling circulation loop are connected with by quick fitting couplings the cooling device “KE”. For this purpose it is necessary to move back the coupling’s bush, insert the hose and release the coupling. It fixes the hose (without screwing or rotation) by groove. Check by easy pull if the hoses’ connections are firm.

During measurements in the temperature range below -10°C and above +90°C the “FTK-CC” / “ME- CP/PP“ may be put into operation only if the cooling liquid flows through the “KE” in order to prevent overheat of the measuring sensor.

Brookfield Engineering Labs., Inc.

Page 17

M/02-212

Image 17
Contents Brookfield R/S Rheometer Brookfield Engineering Labs., Inc 02-212 Contents Appendix Measuring Principle General DescriptionUse of the Rheometer II.1 R/S Rheometer II. System ConfigurationRheometer main instrument minimum configuration II.3 Computer System II.2 Measuring DevicesIII.1 Stand Mounting III. Instrument InstallationHints for height adjustment of the stand III.2 Electrical Connections III.2.1 Temperature Sensor Pt100III.2.3 Printer Connection III.2.2 AC-AdaptorIII.2.4 Computer Connection Use data link cable of Brookfield delivery onlyFTK-CC III.3 Mounting of temperature controlled measuring devicesThermostat connection to temperature control device FTK-CC III.3.2 Mounting of the measuring device ME-CP/PP Thermostat connection to temperature control device FTK-CCBrookfield Engineering Labs., Inc 02-212 Brookfield Engineering Labs., Inc 02-212 Mounting III.3.3 Mounting of the cooling device KEIV.3 Cleaning IV. Environment, handling, cleaning and maintenanceIV.1 Operating environment, Storage IV.2 HandlingIV.4 Maintenance Measurement directly in the substance Measuring SystemsDepth of immersion Measurement with substance in measuring systemMeasurement with temperature control device FTK-CC Turn left Turn right Lower Lift Brookfield Engineering Labs., Inc 02-212 Voltage check VCC15.25 VI. Operation and Menue SystemRheometer Ver .02 #xxxxxx 07.05.98 Brookfield Brookfield VI.1 KeyboardKey-function dependent to executed operation Zero-point calibration VI.2 Menu System of R/S RheometerDefine/edit program Print programs from memoryMenu handling CONFIGURATION-,emie Set time/date Set RS232 Param LanguageVI.3 Selecting from Lists Main-menue Run Single Run Program RemoteSelect Measuring system CC48 CC45 VI.4 Input of numerical values and alphanumeric textsInput Values Val.1/s0100.00 Nr. of MP010 Times Input Values Val.1/s0200.00 Nr. of MP010 Times0100Input Values Val.1/s0290.00 Nr. of MP Times0100 Input Values Val.1/s0200.00 Nr. of MP Times0100VI.5.1 MAIN-Menue → Run Single VI.5 Menue entries Main menueProg. Running Wait for 1.MP Output of MPs to No output device Memory Menu ST StartVI.5.2 MAIN-Menue → Run Program Program end Single mode Total time100s Total #MP10No valid program Enter program first Select program Prog xyz Test Prog ÖlProg. running Wait for 1.MP #1 Output of MPs to Printer Memory MenueVI.5.3 Main menue → Remote VI.5.5 Main menue → ConfigurationVI.5.4 Main menue → Utilities Program end Test Total time200s Total #MP40VI.6.2 Utilities → Edit Programs VI.6 Menue entries in the Utilities-MenueVI.6.1 Utilities → Zero Calibration D1/s=ft Select Meas. system CC45 CC25 CC14 Select Prog to edit Prog xyz Test NEWNr. of steps Select Input-Mode -D1/s Nrpm TauPaEdit Program name Range-Error Range-Error Start-Value min 0.90 maxInput Step Nr time s VI.6.4 Utilities → Measuring systems VI.6.3 Utilities → Print ProgramsTest Press OK to save Printer not ready OK cont. ST stopEnter Meas. Sys. #4 Name CP25 tauprom 01.1418 kgamma VI.6.5 Utilities → Print memoryPress OK to save Name CC25 tauprom 01.1418 kgamma Sel. Output-device Printer RS232VI.6.6 Utilities → Clear memory VI.6.7 Utilities → Measure temperatureVI.7.1 Configuration → Set output mode VI.7 Menu entries of the Configuration menuService, Service2 for service personnel only Select output device No output device Printer RS232VI.7.4 Configuration → Set Time/Date VI.7.2 Configuration → MeasCount modeVI.7.3 Configuration → Reset meascount RS232 4800,8,n,1 Change RS232-Par.? Okyes Stno VI.7.5 Configuration → Set RS232 ParameterNew Time/Date 1715 Press OK to save VI.7.7 Configuration → Service, Service VI.7.6 Configuration → LanguageVI.8 Serial Data transfer via Interface RS232 Baudrate -4800 9600Brookfield Engineering Labs., Inc 02-212 Brookfield Engineering Labs., Inc 02-212 VII. Measurements VII.1 Measuring in manual modeMeasurement control Measurement with FTK-CCVII.2 Measuring in Remote-mode Measurement in remote control modeVIII. Technical data Measured or evaluated values Weight Temperature rangeIX. Guarantee Given ranges are standard values not maximum values A1Data sheets of standard measuring systemsDG DIN Table Plate/Plate measuring systems DIN Measuring system C25-1 C25-2 C50-1 C50-2 C75-1 C75-2Range Error A2 Error messagesPrinter Error Range error End value min 0.90 maxBreak torque max! Single mode Total MP Total time Zero Calibration ErrorError #3 Zero cal. error Please retry cal ST stop Break Speed MAX Break Speed max! Single mode Total MP Total timePC Side A3 Pin-layout of the serial data cableRheometer side General information A4 Requirements to the mains connecting cablesUSA and Canada Other countriesEnglishGermanEnglishGerman A5 Language cross referenceGerman menue system of RC20 Index

M/02-212 specifications

The IBM M/02-212 is a cutting-edge platform that showcases the company’s dedication to excellence in technology and innovation. Designed for both enterprise performance and flexibility, this model adheres to IBM's high standards for processing power and reliability, making it a standout option for businesses looking to enhance their operations.

One of the main features of the M/02-212 is its powerful multi-core processor, which provides exceptional speed and efficiency. This architecture enables the system to handle multiple tasks simultaneously without compromising performance, making it ideal for data-intensive applications and complex processing tasks. It supports a range of workloads, from traditional enterprise applications to modern cloud-based solutions.

In terms of memory, the M/02-212 offers extensive capacity options, utilizing high-speed RAM technology that ensures swift data access and reduced latency. The combination of powerful processing and ample memory resources allows organizations to execute demanding applications and large-scale databases seamlessly.

The system is designed with advanced security features that are integral for protecting sensitive information. Encryption capabilities, secure boot options, and comprehensive identity management functionalities are built-in, providing a robust defense against potential cyber threats. This is particularly valuable for industries like finance and healthcare, where data integrity is paramount.

Moreover, the IBM M/02-212 leverages state-of-the-art virtualization technologies, enabling companies to optimize resource utilization and increase operational efficiency. Through virtualization, multiple virtual machines can run on a single physical server, allowing businesses to reduce hardware costs and improve overall system management.

The M/02-212 also emphasizes scalability, which means organizations can easily expand their capabilities as their needs evolve. This flexibility allows businesses to adapt to changing market conditions without requiring a complete overhaul of their existing infrastructure.

Additionally, it comes equipped with a user-friendly management interface, providing administrators with the tools necessary to monitor system performance, allocate resources, and troubleshoot issues efficiently.

In summary, the IBM M/02-212 combines powerful processing capabilities, robust security measures, advanced virtualization, and exceptional scalability. These characteristics make it an excellent choice for enterprises looking to enhance their IT infrastructure while ensuring efficiency and data security.