IBM M/02-212 Output of MPs to Printer Memory Menue, Prog. running Wait for 1.MP #1

Page 35

If you have passed these points, before starting a measurement, the Rheometer will indicate, where the measure points will be written to.

Output of MPs to:

-Printer

 

-Memory

 

<^> Menue

<ST> Start

At this point you either can start a measurement by pressing

using key.

ST

key, or return back to Main menue

In this example we see, that measurement points will be saved in the memory and will be printed at the printer, so be sure that printer is connected to the Rheometer and ready for operation (=output device). Output devices are printer or serial interface RS232 of the Rheometer or no device (no output device). Their preliminary selection is described in Chapter “Configuration Set Output Mode”.

In case, the Memory is full with measurement data, you have to cancel function Run Single, print out the measurement data from the memory or send data to a PC via serial interface (see Chapter “Utilities Print Memory”) and then delete measurement data from the Memory (see Chapter “Utilities Delete Memory”).

However, you can start a measurement even with full Memory, in this case measurement results will be shown only at LCD display and will be lost after completion of the measurement (because no output device archives measured data).

If you start the measurement, the instrument shows:

Prog. running...

Wait for 1.MP #1

This message will be displayed until the first measurement point is calculated.

First Measure point is displayed like this:

Torque

 

 

 

 

 

 

 

 

100.5 ‰

000.90 1/s

 

Eta:

 

1.728 Pas

Duration

Tau:

 

1.572 Pa

 

05s

20.7°C

#01

 

Temperature

Actual value of preset parameter

Measurment result

Step (in Run Single always 01)

If torque is less than 10 ‰ measurement results are often of low precision. In this case, preset values are to be changed in such a manner, that the torque is higher than 10 ‰.

If the display field for the torque indicates: “M low!”, these values are outside the range of resolution of the Rheometer.

If the temperature is not to displayed, the measuring sensor is not connected. At printout, the temperature is printed as 1000.0°C!

The display is updated on every new measure point. Current measurement can be canceled at any time by pushing ST key.

Brookfield Engineering Labs., Inc.

Page 35

M/02-212

Image 35
Contents Brookfield R/S Rheometer Brookfield Engineering Labs., Inc 02-212 Contents Appendix Measuring Principle General DescriptionUse of the Rheometer II.1 R/S Rheometer II. System ConfigurationRheometer main instrument minimum configuration II.3 Computer System II.2 Measuring DevicesIII.1 Stand Mounting III. Instrument InstallationHints for height adjustment of the stand III.2 Electrical Connections III.2.1 Temperature Sensor Pt100Use data link cable of Brookfield delivery only III.2.2 AC-AdaptorIII.2.3 Printer Connection III.2.4 Computer ConnectionFTK-CC III.3 Mounting of temperature controlled measuring devicesThermostat connection to temperature control device FTK-CC III.3.2 Mounting of the measuring device ME-CP/PP Thermostat connection to temperature control device FTK-CCBrookfield Engineering Labs., Inc 02-212 Brookfield Engineering Labs., Inc 02-212 Mounting III.3.3 Mounting of the cooling device KEIV.2 Handling IV. Environment, handling, cleaning and maintenanceIV.3 Cleaning IV.1 Operating environment, StorageIV.4 Maintenance Measurement directly in the substance Measuring SystemsDepth of immersion Measurement with substance in measuring systemMeasurement with temperature control device FTK-CC Turn left Turn right Lower Lift Brookfield Engineering Labs., Inc 02-212 Voltage check VCC15.25 VI. Operation and Menue SystemRheometer Ver .02 #xxxxxx 07.05.98 Brookfield Brookfield VI.1 KeyboardKey-function dependent to executed operation Print programs from memory VI.2 Menu System of R/S RheometerZero-point calibration Define/edit programMain-menue Run Single Run Program Remote CONFIGURATION-,emie Set time/date Set RS232 Param LanguageMenu handling VI.3 Selecting from ListsInput Values Val.1/s0200.00 Nr. of MP010 Times0100 VI.4 Input of numerical values and alphanumeric textsSelect Measuring system CC48 CC45 Input Values Val.1/s0100.00 Nr. of MP010 TimesInput Values Val.1/s0290.00 Nr. of MP Times0100 Input Values Val.1/s0200.00 Nr. of MP Times0100VI.5.1 MAIN-Menue → Run Single VI.5 Menue entries Main menueProg. Running Wait for 1.MP Output of MPs to No output device Memory Menu ST StartSelect program Prog xyz Test Prog Öl Program end Single mode Total time100s Total #MP10VI.5.2 MAIN-Menue → Run Program No valid program Enter program firstProg. running Wait for 1.MP #1 Output of MPs to Printer Memory MenueProgram end Test Total time200s Total #MP40 VI.5.5 Main menue → ConfigurationVI.5.3 Main menue → Remote VI.5.4 Main menue → UtilitiesVI.6.2 Utilities → Edit Programs VI.6 Menue entries in the Utilities-MenueVI.6.1 Utilities → Zero Calibration D1/s=ft Select Input-Mode -D1/s Nrpm TauPa Select Prog to edit Prog xyz Test NEWSelect Meas. system CC45 CC25 CC14 Nr. of stepsEdit Program name Range-Error Range-Error Start-Value min 0.90 maxInput Step Nr time s Printer not ready OK cont. ST stop VI.6.3 Utilities → Print ProgramsVI.6.4 Utilities → Measuring systems Test Press OK to saveSel. Output-device Printer RS232 VI.6.5 Utilities → Print memoryEnter Meas. Sys. #4 Name CP25 tauprom 01.1418 kgamma Press OK to save Name CC25 tauprom 01.1418 kgammaVI.6.6 Utilities → Clear memory VI.6.7 Utilities → Measure temperatureSelect output device No output device Printer RS232 VI.7 Menu entries of the Configuration menuVI.7.1 Configuration → Set output mode Service, Service2 for service personnel onlyVI.7.4 Configuration → Set Time/Date VI.7.2 Configuration → MeasCount modeVI.7.3 Configuration → Reset meascount RS232 4800,8,n,1 Change RS232-Par.? Okyes Stno VI.7.5 Configuration → Set RS232 ParameterNew Time/Date 1715 Press OK to save Baudrate -4800 9600 VI.7.6 Configuration → LanguageVI.7.7 Configuration → Service, Service VI.8 Serial Data transfer via Interface RS232Brookfield Engineering Labs., Inc 02-212 Brookfield Engineering Labs., Inc 02-212 Measurement with FTK-CC VII.1 Measuring in manual modeVII. Measurements Measurement controlVII.2 Measuring in Remote-mode Measurement in remote control modeVIII. Technical data Measured or evaluated values Weight Temperature rangeIX. Guarantee Given ranges are standard values not maximum values A1Data sheets of standard measuring systemsDG DIN Table Plate/Plate measuring systems DIN Measuring system C25-1 C25-2 C50-1 C50-2 C75-1 C75-2Range error End value min 0.90 max A2 Error messagesRange Error Printer ErrorBreak torque max! Single mode Total MP Total time Zero Calibration ErrorError #3 Zero cal. error Please retry cal ST stop Break Speed MAX Break Speed max! Single mode Total MP Total timePC Side A3 Pin-layout of the serial data cableRheometer side Other countries A4 Requirements to the mains connecting cablesGeneral information USA and CanadaEnglishGermanEnglishGerman A5 Language cross referenceGerman menue system of RC20 Index

M/02-212 specifications

The IBM M/02-212 is a cutting-edge platform that showcases the company’s dedication to excellence in technology and innovation. Designed for both enterprise performance and flexibility, this model adheres to IBM's high standards for processing power and reliability, making it a standout option for businesses looking to enhance their operations.

One of the main features of the M/02-212 is its powerful multi-core processor, which provides exceptional speed and efficiency. This architecture enables the system to handle multiple tasks simultaneously without compromising performance, making it ideal for data-intensive applications and complex processing tasks. It supports a range of workloads, from traditional enterprise applications to modern cloud-based solutions.

In terms of memory, the M/02-212 offers extensive capacity options, utilizing high-speed RAM technology that ensures swift data access and reduced latency. The combination of powerful processing and ample memory resources allows organizations to execute demanding applications and large-scale databases seamlessly.

The system is designed with advanced security features that are integral for protecting sensitive information. Encryption capabilities, secure boot options, and comprehensive identity management functionalities are built-in, providing a robust defense against potential cyber threats. This is particularly valuable for industries like finance and healthcare, where data integrity is paramount.

Moreover, the IBM M/02-212 leverages state-of-the-art virtualization technologies, enabling companies to optimize resource utilization and increase operational efficiency. Through virtualization, multiple virtual machines can run on a single physical server, allowing businesses to reduce hardware costs and improve overall system management.

The M/02-212 also emphasizes scalability, which means organizations can easily expand their capabilities as their needs evolve. This flexibility allows businesses to adapt to changing market conditions without requiring a complete overhaul of their existing infrastructure.

Additionally, it comes equipped with a user-friendly management interface, providing administrators with the tools necessary to monitor system performance, allocate resources, and troubleshoot issues efficiently.

In summary, the IBM M/02-212 combines powerful processing capabilities, robust security measures, advanced virtualization, and exceptional scalability. These characteristics make it an excellent choice for enterprises looking to enhance their IT infrastructure while ensuring efficiency and data security.